HAND-HELD COMPARED WITH ISOKINETIC DYNAMOMETRY FOR MEASUREMENT OF STATIC KNEE EXTENSION TORQUE (PARALLEL RELIABILITY OF DYNAMOMETERS)

被引:85
作者
BOHANNON, RW
机构
[1] Sch. of Allied Health, Connecticut Univ., Storrs, CT
来源
CLINICAL PHYSICS AND PHYSIOLOGICAL MEASUREMENT | 1990年 / 11卷 / 03期
关键词
D O I
10.1088/0143-0815/11/3/004
中图分类号
Q6 [生物物理学];
学科分类号
071011 ;
摘要
The author compares measurements of static knee extension torque obtained using a hand-held dynamometer (HHD) and an isokinetic dynamometer (IKD). Twenty young healthy women performed two maximum knee extensions efforts against each device. The two HHD measurements did not differ significantly and were highly reliable (intra-class correlation coefficient (ICC)=0.945). The two IKD measurements did not differ significantly and were highly reliable (ICC=0.932). The mean of the two HHD and two IKD measurements did not differ significantly. The inter-instrument reliability was fair (ICC=0.797). Thus, under limited circumstances the instruments may be interchangeable. The less expensive and more portable HHD may be a practical alternative for the clinical measurement of muscle strength.
引用
收藏
页码:217 / 222
页数:6
相关论文
共 16 条
[11]   COMPARISON OF ISOKINETIC MEASUREMENTS WITH TEST REPETITIONS [J].
MAWDSLEY, RH ;
KNAPIK, JJ .
PHYSICAL THERAPY, 1982, 62 (02) :169-172
[12]  
NEWMAN L B, 1949, Arch Phys Med Rehabil, V30, P234
[13]   SHOULDER ABDUCTION FATIGUABILITY [J].
NICKLIN, J ;
KARNI, Y ;
WILES, CM .
JOURNAL OF NEUROLOGY NEUROSURGERY AND PSYCHIATRY, 1987, 50 (04) :423-427
[14]   INTRACLASS CORRELATIONS - USES IN ASSESSING RATER RELIABILITY [J].
SHROUT, PE ;
FLEISS, JL .
PSYCHOLOGICAL BULLETIN, 1979, 86 (02) :420-428
[15]   EVALUATION OF INSTRUMENTS FOR MEASURING GRIP STRENGTH [J].
SOLGAARD, S ;
KRISTIANSEN, B ;
JENSEN, JS .
ACTA ORTHOPAEDICA SCANDINAVICA, 1984, 55 (05) :569-572
[16]  
Vanswearingen J, 1983, J Orthop Sports Phys Ther, V4, P217