HIGH-PERFORMANCE LIQUID-CHROMATOGRAPHIC DETERMINATION OF HYDROGEN-PEROXIDE WITH PEROXYOXALATE CHEMILUMINESCENCE DETECTION

被引:41
作者
NAKASHIMA, K [1 ]
WADA, M [1 ]
KURODA, N [1 ]
AKIYAMA, S [1 ]
IMAI, K [1 ]
机构
[1] UNIV TOKYO,BRANCH HOSP PHARM,BUNKYO KU,TOKYO 112,JAPAN
来源
JOURNAL OF LIQUID CHROMATOGRAPHY | 1994年 / 17卷 / 10期
关键词
D O I
10.1080/10826079408013535
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
A high performance liquid chromatographic method with peroxyoxalate chemiluminescence (PO-CL) detection was developed for the determination of hydrogen peroxide. Hydrogen peroxide separated by a reversed phase was detected with the CL produced by the PO-CL reaction using a mixture of 2,4,6,8-tetrathiomorpholinopyrimido[5,4-d]pyrimidine and bis[2-(3,6,9-trioxadecyloxycarbonyl)-4-nitrophenyl]oxalate or bis(2,4,6-trichlorophenyl)oxalate in acetonitrile as a post column reagent. Hydrogen peroxide could be determined over the range of 5-100 pmol on column and the detection limit was 188 fmol at a signal-to-noise ratio of 3. Relative standard deviation of 10-replicate measurements of peak heights for 50 pmol was 2.36%. The detection system applied to the assay of hydrogen peroxide in cola drinks. Trace amounts of hydrogen peroxide could be determined in six cola drinks. The levels were 68.4+/-2.6-174.2+/-8.4 nmol/dl. Hydrogen peroxide contaminated in commercially available hydroperoxides was also determined by the proposed method. The amounts obtained were 0.061% for t-butyl hydroperoxide and 0.003% for cumene hydroperoxide.
引用
收藏
页码:2111 / 2126
页数:16
相关论文
共 9 条
[1]   IMMOBILIZED FLUOROPHORES IN DYNAMIC CHEMI-LUMINESCENCE DETECTION OF HYDROGEN-PEROXIDE [J].
GUBITZ, G ;
VANZOONEN, P ;
GOOIJER, C ;
VELTHORST, NH ;
FREI, RW .
ANALYTICAL CHEMISTRY, 1985, 57 (11) :2071-2074
[2]   POSTCOLUMN ADJUSTMENT OF CONDITIONS FOR PEROXYOXALATE CHEMILUMINESCENCE DETECTION FOR HIGH-PERFORMANCE LIQUID-CHROMATOGRAPHY [J].
HANAOKA, N .
JOURNAL OF CHROMATOGRAPHY, 1990, 503 (01) :155-165
[3]   APPLICATION OF BIS[4-NITRO-2-(3,6,9-TRIOXADECYLOXYCARBONYL)-PHENYL]OXALATE TO POSTCOLUMN CHEMILUMINESCENCE DETECTION IN HIGH-PERFORMANCE LIQUID-CHROMATOGRAPHY [J].
IMAI, K ;
MATSUNAGA, Y ;
TSUKAMOTO, Y ;
NISHITANI, A .
JOURNAL OF CHROMATOGRAPHY, 1987, 400 :169-176
[4]   STABILITY OF BIS(2,4,6-TRICHLOROPHENYL) OXALATE IN HIGH-PERFORMANCE LIQUID-CHROMATOGRAPHY FOR CHEMI-LUMINESCENCE DETECTION [J].
IMAIZUMI, N ;
HAYAKAWA, K ;
MIYAZAKI, M ;
IMAI, K .
ANALYST, 1989, 114 (02) :161-164
[5]   DETERMINATION OF HYDROGEN-PEROXIDE BY FLOW-INJECTION ANALYSIS WITH ARYL OXALATE-SULFORHODAMINE-101 CHEMILUMINESCENCE [J].
KATAYAMA, M ;
TAKEUCHI, H ;
TANIGUCHI, H .
ANALYTICAL LETTERS, 1991, 24 (06) :1005-1015
[6]  
LATIF MSA, 1988, ANAL CHEM, V60, P2671
[7]   FLOW-INJECTION ANALYSIS WITH CHEMILUMINESCENCE DETECTION OF GLUCOSE AND URIC-ACID USING IMMOBILIZED ENZYME REACTOR [J].
NAKASHIMA, K ;
HAYASHIDA, N ;
KAWAGUCHI, S ;
AKIYAMA, S ;
TSUKAMOTO, Y ;
IMAI, K .
ANALYTICAL SCIENCES, 1991, 7 (05) :715-718
[8]   SYNTHESIS OF PYRIMIDO[5,4-D]PYRIMIDINE DERIVATIVES AND THEIR ULTRAVIOLET-ABSORPTION AND FLUORESCENCE SPECTRAL PROPERTIES [J].
NAKASHIMA, K ;
AKIYAMA, S ;
TSUKAMOTO, Y ;
IMAI, K .
DYES AND PIGMENTS, 1990, 12 (01) :21-26
[9]   PHOTOGRAPHIC DETECTION OF HYDROGEN-PEROXIDE AND GLUCOSE BY PEROXYOXALATE CHEMILUMINESCENCE [J].
NAKASHIMA, K ;
KAWAGUCHI, S ;
GIVENS, RS ;
AKIYAMA, S .
ANALYTICAL SCIENCES, 1990, 6 (06) :833-836