VARIABLE COMPENSATION CHEST RADIOGRAPHY PERFORMED WITH A COMPUTED RADIOGRAPHY SYSTEM - DESIGN CONSIDERATIONS AND INITIAL CLINICAL-EXPERIENCE
被引:6
作者:
DOBBINS, JT
RICE, JJ
GOODMAN, PC
PATZ, EF
RAVIN, CE
机构:
[1] Thoracis Research Division, Department of Radiology, Duke University Medical Center, Durham
来源:
关键词:
IMAGES;
PROCESSING;
LUNG;
RADIOGRAPHY;
MEDIASTINUM;
DIGITAL;
STORAGE PHOSPHOR;
TECHNOLOGY;
D O I:
10.1148/radiology.187.1.8451437
中图分类号:
R8 [特种医学];
R445 [影像诊断学];
学科分类号:
1002 ;
100207 ;
1009 ;
摘要:
The authors describe a variable compensation (VC) technique in which an x-ray equalizer and a computed radiography system are used. The VC technique allows retrospective alteration of equalized chest appearance with maintenance of improved signal-to-noise ratio in dense regions. Two imaging plates are used: one upstream of the patient to record the incident beam profile and one downstream to record the equalized image. Subtraction of a weighted version of the upstream image from the downstream image permits alteration of the appearance of the VC image, from the extremes of simulated-unequalized to highly equalized. VC image appearance was optimized with a real-time workstation. The quality of VC images obtained in 33 patients was evaluated by three chest radiologists. Mediastinal appearance was better on VC equalized images than on conventional screen-film images. The simulation of the appearance of a conventional radiograph with VC proved useful in interpretation of lung appearances on equalized radiographs.