SIMULTANEOUS SPECTROFLUOROMETRIC DETERMINATION OF GLYCEROL AND ETHANOL IN WINE BY FLOW-INJECTION USING IMMOBILIZED ENZYMES

被引:30
作者
MATTOS, IL [1 ]
FERNANDEZROMERO, JM [1 ]
DECASTRO, HDL [1 ]
VALCARCEL, M [1 ]
机构
[1] STATE UNIV CAMPINAS,INST CHEM,CAMPINAS,BRAZIL
关键词
GLYCEROL AND ETHANOL DETERMINATION; IMMOBILIZED ENZYME; FLOW INJECTION; SPECTROFLUOROMETRY; WINE;
D O I
10.1039/an9952000179
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
An automatic spectrofluorimetric method for the simultaneous determination of glycerol and ethanol is presented. The method is based on the use of a flow injection (FI) manifold and dehydrogenase enzyme immobilized on controlled-pore glass. The proposed method allows for the simultaneous determination of glycerol (linear range between 1 and 10 mu g ml(-1), r = 0.9989 and s(r) <1%) and ethanol (linear range between 10 and 100 mu g ml(-1), r = 0.9998 and s(r) <0.9%). It was applied to the determination of both analytes in different wines with recoveries between 97 and 105%. The sampling frequency was 60 samples h(-1). The results obtained on eight wines were compared with those provided by conventional methods and the agreement between them corroborated the usefulness of the proposed method.
引用
收藏
页码:179 / 182
页数:4
相关论文
共 22 条
[21]   FLOW-INJECTION ANALYSIS (FIA) BASED ON ENZYMES OR ANTIBODIES - APPLICATIONS IN THE LIFE SCIENCES [J].
SCHMID, RD ;
KUNNECKE, W .
JOURNAL OF BIOTECHNOLOGY, 1990, 14 (01) :3-31
[22]  
SIM KW, 1991, BIOSENS BIOELECTRON, V16, P317