REUSE OF ENTERAL FEEDING TUBES - A POTENTIAL HAZARD TO THE PATIENT - A STUDY OF THE EFFICIENCY OF A REPRESENTATIVE RANGE OF CLEANING AND DISINFECTION PROCEDURES

被引:5
作者
ANDERTON, A
NWOGUH, CE
机构
[1] Department of Bioscience and Biotechnology, University of Strathclyde, Glasgow, G1 1XW, Royal College Building
关键词
ENTERAL FEEDING SYSTEMS; CLEANING; DISINFECTION; MICROBIAL CONTAMINATION;
D O I
10.1016/0195-6701(91)90157-4
中图分类号
R1 [预防医学、卫生学];
学科分类号
1004 ; 120402 ;
摘要
Some hospitals and manufacturers are now recommending that patients (particularly those on home enteral feeding) remove and re-insert their tubes on a daily basis. This study was carried out to evaluate the effectiveness of a representative range of currently used cleaning procedures in removing bacteria from the lumina of these tubes. One thousand-ml portions of feed experimentally contaminated with 102-103 Klebsiella aerogenes ml-1 were perfused through three types of commonly used polyurethane enteral feeding tubes for 15 h. The tubes were then cleaned by a range of methods including rinsing them with sterile water, sterile water and detergent and/or disinfection with hypochlorite solutions. A further 1000-ml sterile feed was then perfused through the tubes for 15 h and it was found that residual organisms in the tubes multiplied to yield levels of 106-109 colony-forming units (cfu) ml-1 in the feed collected from the distal ends of the tubes after 15 h. It is concluded that none of the cleaning methods tested can be recommended as being totally effective in removing bacteria from the lumina of contaminated tubes. © 1991.
引用
收藏
页码:131 / 138
页数:8
相关论文
共 29 条
[1]  
ADAMS MM, 1984, J AM DIET ASSOC, V84, P68
[2]   DECANTING - A SOURCE OF CONTAMINATION OF ENTERAL FEEDS [J].
ANDERTON, A ;
AIDOO, KE .
CLINICAL NUTRITION, 1990, 9 (03) :157-162
[3]  
ANDERTON A, 1983, HUM NUTR-APPL NUTR, V37, P426
[4]   THE EFFECT OF HANDLING PROCEDURES ON MICROBIAL-CONTAMINATION OF ENTERAL FEEDS [J].
ANDERTON, A ;
AIDOO, KE .
JOURNAL OF HOSPITAL INFECTION, 1988, 11 (04) :364-372
[5]  
ANDERTON A, 1984, CLIN NUTR, V3, P171, DOI 10.1016/0261-5614(84)90002-5
[6]   THE POTENTIAL OF ESCHERICHIA-COLI IN ENTERAL FEEDS TO CAUSE FOOD POISONING - A STUDY UNDER SIMULATED WARD CONDITIONS [J].
ANDERTON, A .
JOURNAL OF HOSPITAL INFECTION, 1984, 5 (02) :155-163
[7]   MICROBIAL ADHESION - A SOURCE OF INFECTION IN PATIENTS FED BY ENTERAL FEEDING [J].
ANDERTON, A .
JOURNAL OF HOSPITAL INFECTION, 1986, 7 (02) :205-208
[8]  
BASTOW MD, 1982, HUM NUTR-APPL NUTR, V36, P213
[9]   NASOGASTRIC HYPERALIMENTATION THROUGH A POLYETHYLENE CATHETER - ALTERNATIVE TO CENTRAL VENOUS HYPERALIMENTATION [J].
BETHEL, RA ;
JANSEN, RD ;
HEYMSFIELD, SB ;
ANSLEY, JD ;
HERSH, T ;
RUDMAN, D .
AMERICAN JOURNAL OF CLINICAL NUTRITION, 1979, 32 (05) :1112-1120
[10]  
BSTOW MD, 1984, CLIN NUTR, V4, P7