SUMMING GE(LI)-COMPTON SPECTROMETER WITH HIGH PEAK-TO-TAIL RATIO

被引:14
作者
HICK, H
PEPELNIK, R
机构
[1] Institute of Physics, Reactor Center Seibersdorf
来源
NUCLEAR INSTRUMENTS & METHODS | 1969年 / 68卷 / 02期
关键词
D O I
10.1016/0029-554X(69)90226-2
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
The operation and performance of a summing Ge(Li)-Ge(Li)-Compton spectrometer are described. The geometry and the electronic adjustment have been optimized with respect to best response function. Peak-to-tail ratios of 400 for 662 keV and 360 for 1120 keV energy have been obtained. The construction of a summing Si(Li)-Ge(Li)-Compton spectrometer is suggested to avoid small residual contributions of the pair production to the response function. © 1969.
引用
收藏
页码:240 / +
页数:1
相关论文
共 11 条
[1]   A SINGLE CRYSTAL GE(LI) COMPTON SPECTROMETER [J].
GRUHN, CR ;
KANE, JV ;
KELLY, WH ;
KUO, T ;
BERZINS, G .
NUCLEAR INSTRUMENTS & METHODS, 1967, 54 (02) :268-&
[2]   A HIGH RESOLUTION COMPTON-SPECTROMETER USING A SEMI-CONDUCTOR DETECTOR [J].
HICK, H ;
RUMPOLD, K ;
WEINZIERL, P .
NUCLEAR INSTRUMENTS & METHODS, 1963, 24 (03) :327-328
[3]  
HIGATSBERGER MJ, 1966, P NUCLEAR MATERIALS, P817
[4]  
HIGATSBERGER MJ, TO BE PUBLISHED
[5]  
HIGATSBERGER MJ, 1964, 3 P INT C PUAE GENF, V9, P577
[6]   A SIMPLE SUMMING COMPTON GE(LI) SPECTROMETER [J].
KANTELE, J ;
SUOMINEN, P .
NUCLEAR INSTRUMENTS & METHODS, 1967, 56 (02) :351-&
[7]   A TOTAL ABSORPTION GE(LI) GAMMA RAY SPECTROMETER [J].
KRANER, HW ;
CHASE, RL .
IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 1968, NS15 (03) :381-+
[8]  
MEYER U, 1968, THESIS 2 PHYS I
[9]   A GE(LI) CONCENTRIC DUODE SPECTROMETER FOR COMPTON SUPPRESSION [J].
PALMS, JM ;
WOOD, RE ;
PUCKETT, OH .
IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 1968, NS15 (03) :397-+
[10]  
PEPELNIK R, 1968, ACTA PHYS AUSTRIACA, V27, P212