MOBILE PHASE VARIATIONS IN THERMOSPRAY LIQUID-CHROMATOGRAPHY MASS-SPECTROMETRY OF PESTICIDES

被引:40
作者
DURAND, G [1 ]
DEBERTRAND, N [1 ]
BARCELO, D [1 ]
机构
[1] CSIC, CID, DEPT ENVIRONM CHEM, C JORDI GIRONA SALGADO 18-26, E-08034 BARCELONA, SPAIN
来源
JOURNAL OF CHROMATOGRAPHY-BIOMEDICAL APPLICATIONS | 1991年 / 562卷 / 1-2期
关键词
D O I
10.1016/0378-4347(91)80603-A
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
The effect of four different mobile phase compositions with reversed-phase methanol-water (50:50) + 0.05 M ammonium acetate, methanol-water (50:50) + 0.05 M ammonium formate, acetonitrile-water (50:50) + 0.05 M ammonium acetate and acetonitrile-water (50:50) + 0.05 M ammonium formate were compared in filament-on thermospray liquid chromatography-mass spectrometry for the determination of carbamate and chlorotriazine pesticides. In the positive-ion mode, [M + H]+ and [M + NH4]+ were generally the base peaks for the chlorotriazines and the carbamates, respectively. Depending on the mobile phase used, other adduct ions obtained corresponded to [M + CH3CN + H]+, [M + CH3OH + NH4]+. [M + CH3COONH4 + NH4 -2H2O]+, [M + CH3CN + NH4]+, [M + CH3COONH4 + H - H2O]+ and the dimer [2M + H]+. In the negative-ion mode, [M - H]- and adducts with the ionizing additive [M + CH3COO]- or [M + HCOO]- were obtained. Other ions for the carbamates carbaryl and oxamyl corresponded to [M - CONHCH3 + CH3COOH]- and [M - CON(CH3)2 + HCOO]-, respectively. The variation of mobile phase composition provides additional structural information in thermospray liquid chromatography-mass spectrometry with no appreciable loss of sensitivity. Applications are reported for the determination of carbamate and chlorotriazine pesticides at the ng/g level in spiked and real soil samples, respectively.
引用
收藏
页码:507 / 523
页数:17
相关论文
共 51 条
  • [31] MAEDER H, 1989, RAPID COMMUN MASS SP, V3, P183
  • [32] HIGH-PERFORMANCE LIQUID-CHROMATOGRAPHY OF S-TRIAZINES AND THEIR DEGRADATION PRODUCTS USING ULTRAVIOLET PHOTOMETRIC AND AMPEROMETRIC DETECTION
    PACAKOVA, V
    STULIK, K
    PRIHODA, M
    [J]. JOURNAL OF CHROMATOGRAPHY, 1988, 442 : 147 - 156
  • [33] HIGH-PERFORMANCE LIQUID-CHROMATOGRAPHY MASS-SPECTROMETRY OF TRIAZINE HERBICIDES
    PARKER, CE
    HANEY, CA
    HARVAN, DJ
    HASS, JR
    [J]. JOURNAL OF CHROMATOGRAPHY, 1982, 242 (01): : 77 - 96
  • [34] GASEOUS ACIDITY AND BASICITY SCALES AS GUIDES TO CHEMICAL IONIZATION IN REVERSED-PHASE LIQUID-CHROMATOGRAPHY MASS-SPECTROMETRY WITH DIRECT LIQUID INTRODUCTION
    PARKER, CE
    BURSEY, MM
    SMITH, RW
    GASKELL, SJ
    [J]. JOURNAL OF CHROMATOGRAPHY, 1985, 347 (01): : 61 - 74
  • [35] DEPENDENCE OF ION FORMATION UPON THE IONIC ADDITIVE IN THERMOSPRAY LIQUID-CHROMATOGRAPHY NEGATIVE-ION MASS-SPECTROMETRY
    PARKER, CE
    SMITH, RW
    GASKELL, SJ
    BURSEY, MM
    [J]. ANALYTICAL CHEMISTRY, 1986, 58 (08) : 1661 - 1664
  • [36] DETERMINATION OF TRACE LEVELS OF HERBICIDES AND THEIR DEGRADATION PRODUCTS IN SURFACE AND GROUND WATERS BY GAS-CHROMATOGRAPHY ION-TRAP MASS-SPECTROMETRY
    PEREIRA, WE
    ROSTAD, CE
    LEIKER, TJ
    [J]. ANALYTICA CHIMICA ACTA, 1990, 228 (01) : 69 - 75
  • [37] RANKIN PC, 1971, J ASSOC OFF ANA CHEM, V54, P1340
  • [38] ROSTAD CE, 1988, 36TH ASMS C MASS SPE, P248
  • [39] RUDEWICZ PJ, 1988, 211 FINN MAT APPL RE, P1
  • [40] THERMOSPRAY MASS-SPECTROMETRY OF N-METHYLCARBAMATES
    SAAR, J