USE OF HEAT AND MOISTURE EXCHANGING (HME) FILTERS IN MECHANICALLY VENTILATED ICU PATIENTS - INFLUENCE ON AIRWAY FLOW-RESISTANCE

被引:30
作者
CHIARANDA, M
VERONA, L
PINAMONTI, O
DOMINIONI, L
MINOJA, G
CONTI, G
机构
[1] UNIV PAVIA, DEPT SURG ANAESTHESIOL & INTENS CARE, VARESE, ITALY
[2] UNIV ROMA LA SAPIENZA, INST ANAESTHESIOL & INTENS CARE, ROME, ITALY
关键词
HEAT; HUMIDITY; RESPIRATION; ARTIFICIAL INSTRUMENTATION; HME;
D O I
10.1007/BF01711088
中图分类号
R4 [临床医学];
学科分类号
1002 ; 100602 ;
摘要
Objective: To investigate the flow-resistance of a new generation of Heat Moisture Exchanging Filters (HME filters) during 24 h of clinical use. Design: Before-after trial. Setting: A general Intensive Care Unit of a university hospital. Patients: A consecutive series of 96 patients undergoing mechanical ventilation for respiratory insufficiency of various etiology and severity. Methods: The characteristics of the secretions collected by tracheal suctioning and the pressure/flow relationship of the HMEs before and after 24 h of clinical use were analyzed. Results: The resistance of the HMEs when dry was 2 hPa/l.s, and it increased to a maximum of 1 hPa/l.s in 83% of the patients after 24 hours; in four patients with particularly heavy secretions HME resistance was 4-5 hPa/l.s. There were no significant modifications of the secretions within the investigation period, excluding, in particular, an increase in density with consequent tracheal tube obstruction. Conclusion: The gas conditioning efficiency and design performance of the tested HMEs did not create a significant obstacle to airflow medium term mechanical ventilation; however, these devices should be cautiously used in patients with heavy bronchial secretions.
引用
收藏
页码:462 / 466
页数:5
相关论文
共 36 条
[1]  
BETHUNE DW, 1985, ANAESTHESIA, V40, P210
[2]  
BOITEAU R, 1990, REAN SOINS INTENS ME, V6, P563
[3]  
BORGHI V, 1992, Acta Anaesthesiologica Italica, V43, P93
[4]  
BOULHAC M, 1984, AGGRESSOLOGIE, V25, P299
[5]   HUMIDIFICATION IN ANESTHESIA - REVIEW OF PRESENT SITUATION [J].
BOYS, JE ;
HOWELLS, TH .
BRITISH JOURNAL OF ANAESTHESIA, 1972, 44 (08) :879-+
[6]   HUMIDITY AND THE ANESTHETIZED PATIENT [J].
CHALON, J ;
PATEL, C ;
ALI, M ;
RAMANATHAN, S ;
CAPAN, L ;
TANG, CK ;
TURNDORF, H .
ANESTHESIOLOGY, 1979, 50 (03) :195-198
[7]  
CHALON J, 1984, ANESTH ANALG, V63, P566
[8]  
CHIARANDA M, 1992, Acta Anaesthesiologica Italica, V43, P73
[9]   ENDOTRACHEAL-TUBE OCCLUSION ASSOCIATED WITH THE USE OF HEAT AND MOISTURE EXCHANGERS IN THE INTENSIVE-CARE UNIT [J].
COHEN, IL ;
WEINBERG, PF ;
FEIN, IA ;
ROWINSKI, GS .
CRITICAL CARE MEDICINE, 1988, 16 (03) :277-279
[10]   EFFECTS OF THE HEAT-MOISTURE EXCHANGERS ON DYNAMIC HYPERINFLATION OF MECHANICALLY VENTILATED COPD PATIENTS [J].
CONTI, G ;
DEBLASI, RA ;
ROCCO, M ;
PELAIA, P ;
ANTONELLI, M ;
BUFI, M ;
MATTIA, C ;
GASPARETTO, A .
INTENSIVE CARE MEDICINE, 1990, 16 (07) :441-443