We have fabricated high-T-c SQUID magnetometers with a pickup coil and a multiturn input coil integrated onto the same substrate. A four-channel system operating in liquid nitrogen has been developed and used to measure biomagnetic signals emanating from both the heart and the brain. The magnetic field sensitivity of the integrated SQUID magnetometer reaches 70 fT/root Hz at 1 kHz and 280 fT/root Hz at 1 Hz. The use of superconductor-normal metal-superconductor (SNS) step-edge junctions with YBa2Cu3O7-delta as the superconductor and Ag-Au alloy as the normal metal allows the measurements to be performed with standard de bias electronics. (C) 1995 American Institute of Physics.
引用
收藏
页码:1926 / 1928
页数:3
相关论文
共 14 条
[1]   DESIGN AND OPTIMIZATION OF DC SQUIDS FABRICATED USING A SIMPLIFIED 4-LEVEL PROCESS [J].
CANTOR, R ;
RYHANEN, T ;
DRUNG, D ;
KOCH, H ;
SEPPA, H .
IEEE TRANSACTIONS ON MAGNETICS, 1991, 27 (02) :2927-2931
[2]  
DiIorio MS, 1993, IEEE T APPL SUPERCON, V3, P2011
[3]   PRACTICAL HIGH TC-JOSEPHSON JUNCTIONS AND DC SQUIDS OPERATING ABOVE 85K [J].
DIIORIO, MS ;
YOSHIZUMI, S ;
YANG, KY ;
ZHANG, J ;
MAUNG, M .
APPLIED PHYSICS LETTERS, 1991, 58 (22) :2552-2554
[4]  
DIIORIO MS, 1993, ADV SUPERCONDUCTIVIT, V5, P1161
[5]   MAGNETOENCEPHALOGRAPHY - THEORY, INSTRUMENTATION, AND APPLICATIONS TO NONINVASIVE STUDIES OF THE WORKING HUMAN BRAIN [J].
HAMALAINEN, M ;
HARI, R ;
ILMONIEMI, RJ ;
KNUUTILA, J ;
LOUNASMAA, OV .
REVIEWS OF MODERN PHYSICS, 1993, 65 (02) :413-497
[6]   PLANAR COUPLING SCHEME FOR ULTRA LOW-NOISE DC SQUIDS [J].
JAYCOX, JM ;
KETCHEN, MB .
IEEE TRANSACTIONS ON MAGNETICS, 1981, 17 (01) :400-403
[7]   DC SQUID MAGNETOMETERS FROM SINGLE LAYERS OF YBA2CU3O7-X [J].
KOELLE, D ;
MIKLICH, AH ;
LUDWIG, F ;
DANTSKER, E ;
NEMETH, DT ;
CLARKE, J .
APPLIED PHYSICS LETTERS, 1993, 63 (16) :2271-2273
[8]   HIGH-T-C MULTILAYER MAGNETOMETERS WITH IMPROVED 1/F NOISE [J].
LUDWIG, F ;
DANTSKER, E ;
KOELLE, D ;
KLEINER, R ;
MIKLICH, AH ;
NEMETH, DT ;
CLARKE, J ;
DRUNG, D ;
KNAPPE, S ;
KOCH, H .
IEEE TRANSACTIONS ON APPLIED SUPERCONDUCTIVITY, 1995, 5 (02) :2919-2922
[9]   INTEGRATED HIGH-T-C MULTILOOP MAGNETOMETER [J].
LUDWIG, F ;
DANTSKER, E ;
KLEINER, R ;
KOELLE, D ;
CLARKE, J ;
KNAPPE, S ;
DRUNG, D ;
KOCH, H ;
ALFORD, NM ;
BUTTON, TW .
APPLIED PHYSICS LETTERS, 1995, 66 (11) :1418-1420
[10]   SENSITIVE YBA2CU3O7-X THIN-FILM MAGNETOMETER [J].
MIKLICH, AH ;
KINGSTON, JJ ;
WELLSTOOD, FC ;
CLARKE, J ;
COLCLOUGH, MS ;
CHAR, K ;
ZAHARCHUK, G .
APPLIED PHYSICS LETTERS, 1991, 59 (08) :988-990