MR IMAGING AND BIOMEDICAL IMPLANTS, MATERIALS, AND DEVICES - AN UPDATED REVIEW

被引:110
作者
SHELLOCK, FG [1 ]
CURTIS, JS [1 ]
机构
[1] UNIV CALIF LOS ANGELES,SCH MED,DEPT DIAGNOST RADIOL,LOS ANGELES,CA 90024
关键词
MAGNETIC RESONANCE (MR); SAFETY;
D O I
10.1148/radiology.180.2.2068325
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Certain ferromagnetic metallic implants, materials, and devices are regarded as contraindications for magnetic resonance imaging, primarily because of the risks associated with their movement or dislodgment. More than 40 publications have reported the ferromagnetic qualities of 261 different metallic objects (aneurysm and hemostatic clips, 32; carotid artery vascular clamps, five; dental devices or materials, 16; heart valve prostheses, 29; intravascular coils, filters, and stents, 14; ocular implants, 12; orthopedic implants, materials, and devices, 15; otologic implants, 56; pellets and bullets, 23; penile implants, nine; vascular access ports, 33; and miscellaneous, 17) on the basis of measurements of deflection forces or attraction during exposure to static magnetic fields at strengths of 0.147-4.7 T. The results of these studies are listed with respect to the specific object tested, the material used to construct the object (if known), whether or not the object was deflected or moved during exposure to the static magnetic field, and the highest static magnetic field strength used for testing the object.
引用
收藏
页码:541 / 550
页数:10
相关论文
共 60 条
  • [1] MAGNETIC-RESONANCE-IMAGING AND RETINAL TACKS
    ALBERT, DW
    OLSON, KR
    PAREL, JM
    HERNANDEZ, E
    LEE, W
    QUENCER, R
    [J]. ARCHIVES OF OPHTHALMOLOGY, 1990, 108 (03) : 320 - 321
  • [2] FURTHER-STUDIES ON THE EFFECTS OF MAGNETIC-RESONANCE-IMAGING FIELDS ON MIDDLE-EAR IMPLANTS
    APPLEBAUM, EL
    VALVASSORI, GE
    [J]. ANNALS OF OTOLOGY RHINOLOGY AND LARYNGOLOGY, 1990, 99 (10) : 801 - 804
  • [3] APPLEBAUM EL, 1985, ARCH OTOLARYNGOL, V111, P820
  • [4] BARRAFATO D, 1984, CAN J SURG, V27, P509
  • [5] BECKER RL, 1988, AM J NEURORADIOL, V9, P885
  • [6] BELLON EM, 1984, MAGN RESON IMAGING, V2, P41
  • [7] MAGNETIC-FIELD EFFECTS ON SURGICAL LIGATION CLIPS
    BROWN, MA
    CARDEN, JA
    COLEMAN, RE
    MCKINNEY, R
    SPICER, LD
    [J]. MAGNETIC RESONANCE IMAGING, 1987, 5 (06) : 443 - 453
  • [8] POTENTIAL HAZARDS IN NMR IMAGING - HEATING EFFECTS OF CHANGING MAGNETIC-FIELDS AND RF FIELDS ON SMALL METALLIC IMPLANTS
    DAVIS, PL
    CROOKS, L
    ARAKAWA, M
    MCREE, R
    KAUFMAN, L
    MARGULIS, AR
    [J]. AMERICAN JOURNAL OF ROENTGENOLOGY, 1981, 137 (04) : 857 - 860
  • [9] INTRAOCULAR-LENS IMPLANTS (PSEUDOPHAKOI) AND STEELWIRE SUTURES - A CONTRAINDICATION FOR MRI
    DEKEIZER, RJW
    STRAKE, LT
    [J]. DOCUMENTA OPHTHALMOLOGICA, 1986, 61 (3-4) : 281 - 284
  • [10] ANEURYSM CLIP MOTION DURING MAGNETIC-RESONANCE IMAGING - INVIVO EXPERIMENTAL-STUDY WITH METALLURGICAL FACTOR-ANALYSIS
    DUJOVNY, M
    KOSSOVSKY, N
    KOSSOWSKY, R
    VALDIVIA, R
    SUK, JS
    DIAZ, FG
    BERMAN, SK
    CLEARY, W
    [J]. NEUROSURGERY, 1985, 17 (04) : 543 - 548