OPTIMIZATION AND USE OF FLOW-INJECTION VAPOR GENERATION INDUCTIVELY COUPLED PLASMA MASS-SPECTROMETRY FOR THE DETERMINATION OF ARSENIC, ANTIMONY AND MERCURY IN WATER AND SEA-WATER AT ULTRATRACE LEVELS

被引:69
作者
STROH, A
VOLLKOPF, U
机构
关键词
FLOW INJECTION INDUCTIVELY COUPLED PLASMA MASS SPECTROMETRY; ONLINE VAPOR GENERATION; OPTIMIZATION OF FLOW INJECTION PARAMETERS; MATRIX EFFECTS; CERTIFIED REFERENCE MATERIALS FOR WATER;
D O I
10.1039/ja9930800035
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
Vapour generation flow injection inductively coupled plasma mass spectrometry (FI-ICP-MS) was used for the determination of As, Sb and Hg in four international water reference materials (National Research Council of Canada). A commercially available FI device was easily connected to the ICP-MS system. Flow injection parameters such as sample volume, purge gas flow rate and concentration of reductant were investigated and optimized (univariate) as were different sample pre-reduction techniques. The precision and accuracy of the results obtained show the applicability of this method to the determination of vapour-forming elements at ultratrace levels in environmental samples. Detection limits are in the range 0.5-7 ng l-1 for Bi, Sb, Se-82, Te, Hg and As (with pre-reduction).
引用
收藏
页码:35 / 40
页数:6
相关论文
共 34 条
[21]   SAMPLE ANALYSIS USING PLASMA SOURCE-MASS SPECTROMETRY WITH ELECTROTHERMAL SAMPLE INTRODUCTION [J].
PARK, CJ ;
VANLOON, JC ;
ARROWSMITH, P ;
FRENCH, JB .
ANALYTICAL CHEMISTRY, 1987, 59 (17) :2191-2196
[22]   FLOW INJECTION ANALYSES .1. NEW CONCEPT OF FAST CONTINUOUS-FLOW ANALYSIS [J].
RUZICKA, J ;
HANSEN, EH .
ANALYTICA CHIMICA ACTA, 1975, 78 (01) :145-157
[23]   SORBENT EXTRACTION IN FLOW-INJECTION ANALYSIS AND ITS APPLICATION TO ENHANCEMENT OF ATOMIC SPECTROMETRY [J].
RUZICKA, J ;
ARNDAL, A .
ANALYTICA CHIMICA ACTA, 1989, 216 (1-2) :243-255
[24]  
RUZICKA J, 1988, ANAL CHIM ACTA, V180, P1
[25]  
Ruzicka J., 1981, FLOW INJECTION ANAL
[26]  
SCHRADER W, 1989, LABORPRAXIS, V5, P406
[27]   DETERMINATION OF ARSENIC, ANTIMONY, BISMUTH, SELENIUM AND TIN IN BIOLOGICAL AND ENVIRONMENTAL-SAMPLES BY CONTINUOUS-FLOW HYDRIDE GENERATION ICP-AES WITHOUT GAS-LIQUID SEPARATOR [J].
SCHRAMEL, P ;
XU, LQ .
FRESENIUS JOURNAL OF ANALYTICAL CHEMISTRY, 1991, 340 (01) :41-47
[28]   FLOW-INJECTION ANALYSIS TECHNIQUES FOR ATOMIC-ABSORPTION SPECTROMETRY - A REVIEW [J].
TYSON, JF .
ANALYST, 1985, 110 (05) :419-429
[29]  
VOLLKOPF U, 1990, ATOM SPECTROSC, V11, P135
[30]  
Vollkopf U., 1992, FRESEN J ANAL CHEM, V342, P917