DISPOSABLE, SHEATHED, FLEXIBLE SIGMOIDOSCOPY - A PROSPECTIVE, MULTICENTER, RANDOMIZED TRIAL

被引:32
作者
ROTHSTEIN, RI [1 ]
LITTENBERG, B [1 ]
机构
[1] DARTMOUTH HITCHCOCK MED CTR,DEPT MED,TECHNOL ASSESSMENT PROGRAM,LEBANON,NH 03756
关键词
D O I
10.1016/S0016-5107(95)70192-3
中图分类号
R57 [消化系及腹部疾病];
学科分类号
摘要
A new fiberoptic sigmoidoscopic system has been developed that utilizes a disposable sheath to cover and protect all working surfaces of the endoscope from contamination. The reusable part of the endoscope has no air, water, or suction/biopsy channels. These are incorporated in the disposable sheath, which is easily removed after use to provide each patient with a contamination-free endoscope. A prospective, randomized, controlled trial was performed to compare the disposable, sheathed, flexible sigmoidoscope with standard sigmoidoscopes. Clinical evaluations of the new sigmoidoscope system were performed at 15 facilities. Visual analog rating scales were used to record evaluations of endoscope performance and reprocessing by endoscopists and reprocessing personnel. The time to perform procedures, depth of insertion, and total instrument downtime were also recorded. One hundred forty-three procedures (70 standard, 73 sheathed) were performed. No significant difference was found for overall depth of insertion (50 versus 48 cm), although fewer sheathed endoscopes reached to 60 cm than did standard endoscopes (51% versus 30%). The sheathed system had a slightly longer mean procedure time than the standard (5.6 versus 6.7 minutes), but a significantly shortened overall downtime (32.8 vs 8.1 minutes). The standard system was preferred by the endoscopists. Reprocessing personnel preferred the disposable system. The disposable sigmoidoscope system has important advantages of decreased instrument turn-around time and potentially increased staff and patient safety, and future models should be improved to meet physicians' concerns.
引用
收藏
页码:566 / 572
页数:7
相关论文
共 34 条
[1]   LONG-TERM RISK OF COLORECTAL-CANCER AFTER EXCISION OF RECTOSIGMOID ADENOMAS [J].
ATKIN, WS ;
MORSON, BC ;
CUZICK, J .
NEW ENGLAND JOURNAL OF MEDICINE, 1992, 326 (10) :658-662
[2]   DISINFECTION AND ENDOSCOPY - SUMMARY AND RECOMMENDATIONS [J].
AXON, ATR .
JOURNAL OF GASTROENTEROLOGY AND HEPATOLOGY, 1991, 6 (01) :23-24
[3]  
AXON ATR, 1981, LANCET, V1, P1093
[4]   HYDROGEN-PEROXIDE ENTERITIS - THE SNOW WHITE SIGN [J].
BILOTTA, JJ ;
WAYE, JD .
GASTROINTESTINAL ENDOSCOPY, 1989, 35 (05) :428-430
[5]  
BOND WW, 1991, DISINFECTION STERILI, P1097
[6]   HOW WELL ARE ENDOSCOPES CLEANED AND DISINFECTED BETWEEN PATIENTS [J].
COLLIGNON, P ;
GRAHAM, E .
MEDICAL JOURNAL OF AUSTRALIA, 1989, 151 (05) :269-+
[7]   ASTHMA AND RHINITIS AFTER EXPOSURE TO GLUTARALDEHYDE IN ENDOSCOPY UNITS [J].
CORRADO, OJ ;
OSMAN, J ;
DAVIES, RJ .
HUMAN TOXICOLOGY, 1986, 5 (05) :325-327
[8]  
CWALINA WM, 1988, SOC GASTROINTESTINAL, P100
[9]   INVESTIGATION OF AN OUTBREAK OF BLOODY DIARRHEA - ASSOCIATION WITH ENDOSCOPIC CLEANING SOLUTION AND DEMONSTRATION OF LESIONS IN AN ANIMAL-MODEL [J].
DURANTE, L ;
ZULTY, JC ;
ISRAEL, E ;
POWERS, PJ ;
RUSSELL, RG ;
QIZILBASH, AH ;
MORRIS, JG .
AMERICAN JOURNAL OF MEDICINE, 1992, 92 (05) :476-480
[10]  
FAVERO MS, 1991, INFECT CONT HOSP EP, V12, P279