Several variables that can affect reproducibility of both the peak height and the migration time in capillary electrophoresis (CE) were investigated here. A great part of the imprecision in CE for migration time is related to the interaction of the analyte with the capillary wall. Within a run, acidic compounds with long migration times have a higher relative standard deviation. In general, conditions that decreased the migration time, such as a short capillary length, tended to enhance the reproducibility of the migration time. Sample size also affected the reproducibility of peak height. Too small or too large an injection volume diminished the reproducibility. Sample matrix, such as high levels of proteins or salts in the sample, decreased the reproducibility. Furthermore, improvement in reproducibility can be achieved by using internal standards (especially for migration time), thorough washing of the capillary, or removing the excess proteins by acetonitrile deproteinization. Acetonitrile has an added effect of producing sample stacking. Depending on the migration time and analyte concentration, the peak height and area yielded different reproducibility results.
引用
收藏
页码:2159 / 2163
页数:5
相关论文
共 13 条
[1]   QUANTITATIVE APPLICATIONS AND VALIDATION OF THE RESOLUTION OF ENANTIOMERS BY CAPILLARY ELECTROPHORESIS [J].
ALTRIA, KD ;
GOODALL, DM ;
ROGAN, MM .
ELECTROPHORESIS, 1994, 15 (06) :824-827
[2]   USE OF HIGH IONIC-STRENGTH BUFFERS FOR THE SEPARATION OF PROTEINS AND PEPTIDES WITH CAPILLARY ELECTROPHORESIS [J].
CHEN, FA ;
KELLY, L ;
PALMIERI, R ;
BIEHLER, R ;
SCHWARTZ, H .
JOURNAL OF LIQUID CHROMATOGRAPHY, 1992, 15 (6-7) :1143-1161
[3]   INTERNAL STANDARDIZATION TECHNIQUE FOR CAPILLARY ZONE ELECTROPHORESIS [J].
DOSE, EV ;
GUIOCHON, GA .
ANALYTICAL CHEMISTRY, 1991, 63 (11) :1154-1158
[4]   SAMPLE MATRIX EFFECTS IN CAPILLARY ELECTROPHORESIS .1. BASIC CONSIDERATIONS [J].
GARCIA, LL ;
SHIHABI, ZK .
JOURNAL OF CHROMATOGRAPHY A, 1993, 652 (02) :465-469
[5]   ANALYSIS OF FACTORS CAUSING PEAK BROADENING IN CAPILLARY ZONE ELECTROPHORESIS [J].
HUANG, XH ;
COLEMAN, WF ;
ZARE, RN .
JOURNAL OF CHROMATOGRAPHY, 1989, 480 :95-110
[6]   SAMPLE MATRIX EFFECTS IN CAPILLARY ELECTROPHORESIS .2. ACETONITRILE DEPROTEINIZATION [J].
SHIHABI, ZK .
JOURNAL OF CHROMATOGRAPHY A, 1993, 652 (02) :471-475
[7]  
SHIHABI ZK, IN PRESS J CAP ELECT
[8]  
SHIHABI ZK, 1994, HDB ELECTROPHORESIS, P537
[9]   INTRODUCTION OF MIGRATION INDEXES FOR IDENTIFICATION - CHIRAL SEPARATION OF SOME BETA-BLOCKERS BY USING CYCLODEXTRINS IN MICELLAR ELECTROKINETIC CAPILLARY CHROMATOGRAPHY [J].
SIREN, H ;
JUMPPANEN, JH ;
MANNINEN, K ;
RIEKKOLA, ML .
ELECTROPHORESIS, 1994, 15 (06) :779-784
[10]   INFLUENCE OF OPERATING PARAMETERS ON REPRODUCIBILITY IN CAPILLARY ELECTROPHORESIS [J].
SMITH, SC ;
STRASTERS, JK ;
KHALEDI, MG .
JOURNAL OF CHROMATOGRAPHY, 1991, 559 (1-2) :57-68