AUTOMATED SINGLE-VOXEL PROTON MRS - TECHNICAL DEVELOPMENT AND MULTISITE VERIFICATION

被引:224
作者
WEBB, PG
SAILASUTA, N
KOHLER, SJ
RAIDY, T
MOATS, RA
HURD, RE
机构
[1] GE CO,MED SYST,47697 WESTINGHOUSE DR,FREMONT,CA 94539
[2] HUNTINGTON MED RES INST,PASADENA,CA
关键词
PROTON MRS; STEAM; AUTOMATION; MULTISITE TRIAL;
D O I
10.1002/mrm.1910310404
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
To improve clinical utility, an integrated method has been developed to automatically acquire and process single-voxel in vivo proton spectra on a 1.5 T clinical scanner. This method includes automated adjustment of linear shims using a very rapid modified simplex method, automated water suppression, and applies a water referencing scheme to correct for phase and residual eddy current effects. No operator intervention is required for the acquisition and processing of these pure-absorption spectra. This method was tested in a preliminary multisite trial to determine intersite and intrasite variability of metabolite ratio measurements. In a sample of over 100 examinations, the standard deviation of the ratios NAA:Cr, Cho:Cr, and ml:Cr were found to be under 15% when using this method, a substantially narrower range than has been found in studies relying on manual adjustment of the instrument and/or manual processing. This result indicates that automated setting of acquisition and processing parameters is of critical importance in the clinical application of in vivo spectroscopy.