AUTOMATED SAMPLE PREPARATION USING SUPPORTED LIQUID MEMBRANES FOR LIQUID-CHROMATOGRAPHIC DETERMINATION OF SULFONYLUREA HERBICIDES

被引:35
作者
NILVE, G
STEBBINS, R
机构
[1] Analytical Chemistry, University of Lund, Lund, S-221 00
关键词
COLUMN LIQUID CHROMATOGRAPHY; LIQUID MEMBRANE SAMPLE PREPARATION; SULFONYLUREA HERBICIDES; UV DETECTION; PHOTOLYSIS ELECTROCHEMICAL DETECTION; CHLORSULFURON; TRANSPORT; PRECONCENTRATION; SEPARATIONS; RESIDUES; CLEANUP; AMINES; PHASE;
D O I
10.1007/BF02276252
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
Sample preparation for determination of sulfonylurea herbicides in aqueous samples is investigated. The technique studied utilizes extraction and back extraction in an automated flow system and is coupled on-line to a liquid chromatographic system. The extraction unit consists of an immobilized liquid membrane, separating two aqueous phases. From the acidified donor phase the analytes are extracted into the organic solvent of the membrane. After traversing the membrane they are back extracted into an alkaline/neutral aqueous acceptor phase. They are trapped in the acceptor by dissociation, making them insoluble in the membrane. Studies of the sample preparation system concern factors like channel length of separators, distribution coefficients of analytes and use of a precolumn instead of loop for chromatographic injections. Effects of the internal diameter of the analytical column as well as the detection of the sulfonylureas are investigated.
引用
收藏
页码:269 / 277
页数:9
相关论文