MICROSEQUENCE ANALYSIS OF PEPTIDES AND PROTEINS .9. AN IMPROVED, COMPACT, AUTOMATED INSTRUMENT

被引:14
作者
CALAYCAY, J [1 ]
RUSNAK, M [1 ]
SHIVELY, JE [1 ]
机构
[1] CITY HOPE NATL MED CTR,BIOMED INSTRUMENTAT SERV,DUARTE,CA 91010
关键词
D O I
10.1016/0003-2697(91)90177-U
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
We describe the construction of an improved, compact protein sequencer with a vertical flow path and continuous flow reactor (CFR). Unique features include a hexagonal valve for six fluid inputs to the CFR, which connects vertically to a transfer valve that allows sample, reagent, and solvent input to a conversion flask (CF). The simplified CF contains only two inputs at the top, one for sample, reagent, and solvent input, and the other a vent. The CF drains from the bottom, connecting to a switching valve which allows either delivery to waste or to an on-line HPLC for the analysis of phenylthiohydantoin amino acid derivatives. Approximately 90% of the sample is analyzed by use of a sonic flow detector. The overall vertical flow path of the sequencer is about 16 cm. The size of the instrument (25 w × 38 h × 44 d cm) is smaller than that of commercially available sequencers or HPLC systems. The performance of the instrument includes reduced background peaks and high-sensitivity sequence analysis at the 5-10 pmol level. The simplified sequencer is more economical and portable than conventional sequencers. © 1991.
引用
收藏
页码:23 / 31
页数:9
相关论文
共 19 条
[1]   A PROTEIN SEQUENATOR [J].
EDMAN, P ;
BEGG, G .
EUROPEAN JOURNAL OF BIOCHEMISTRY, 1967, 1 (01) :80-&
[2]  
FISCHER S, 1990, 1990 PITTSB C
[3]  
HAWKE D, 1982, ANAL BIOCHEM, V120, P302, DOI 10.1016/0003-2697(82)90351-7
[4]   MICROSEQUENCE ANALYSIS OF PEPTIDES AND PROTEINS .5. DESIGN AND PERFORMANCE OF A NOVEL GAS-LIQUID SOLID-PHASE INSTRUMENT [J].
HAWKE, DH ;
HARRIS, DC ;
SHIVELY, JE .
ANALYTICAL BIOCHEMISTRY, 1985, 147 (02) :315-330
[5]  
HEWICK RM, 1981, J BIOL CHEM, V256, P7990
[6]   DIRECT MICROSEQUENCE ANALYSIS OF POLYPEPTIDES USING AN IMPROVED SEQUENATOR, A NON-PROTEIN CARRIER (POLYBRENE), AND HIGH-PRESSURE LIQUID-CHROMATOGRAPHY [J].
HUNKAPILLER, MW ;
HOOD, LE .
BIOCHEMISTRY, 1978, 17 (11) :2124-2133
[7]  
KENT S, 1987, BIOTECHNIQUES, V5, P314
[8]  
KLAPPER DG, 1978, ANAL BIOCHEM, V85, P126, DOI 10.1016/0003-2697(78)90282-8
[9]   SOLID-PHASE EDMAN DEGRADATION - AUTOMATIC PEPTIDE SEQUENCER [J].
LAURSEN, RA .
EUROPEAN JOURNAL OF BIOCHEMISTRY, 1971, 20 (01) :89-&
[10]   SOLID-PHASE SEQUENCE-ANALYSIS OF PROTEINS ELECTROBLOTTED OR SPOTTED ONTO POLYVINYLIDENE DIFLUORIDE MEMBRANES [J].
PAPPIN, DJC ;
COULL, JM ;
KOSTER, H .
ANALYTICAL BIOCHEMISTRY, 1990, 187 (01) :10-19