VARIABILITY OF ARTERIAL BLOOD-GAS VALUES OVER TIME IN STABLE MEDICAL ICU PATIENTS

被引:36
作者
SASSE, SA
CHEN, PA
MAHUTTE, CK
机构
[1] VET AFFAIRS MED CTR,DEPT MED,PULM SECT 111P,LONG BEACH,CA 90822
[2] UNIV CALIF IRVINE,IRVINE,CA
关键词
ARTERIAL BLOOD GASES; FLUORESCENT SENSORS; INTENSIVE CARE; MONITORING;
D O I
10.1378/chest.106.1.187
中图分类号
R4 [临床医学];
学科分类号
1002 ; 100602 ;
摘要
The spontaneous variability of arterial blood gas and pH values (ABGs) was examined in a group of 28 typical stable medical ICU patients under a variety of ventilatory conditions, In each patient, 13 ABG specimens were measured at 5-min intervals during a 1-h study period using a new bedside, extravascular fluorescent blood gas monitor. For all patients, the mean coefficient of variation (C) was 6.1 percent for PO2 and 4.7 percent for PCO2 The average SD for pH was 0.012. We conclude that the spontaneous variability for ABG values over a 1-h period is substantial and that this variability should be taken into account when making clinical decisions based on ABG values.
引用
收藏
页码:187 / 193
页数:7
相关论文
共 18 条
[1]   STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENT [J].
BLAND, JM ;
ALTMAN, DG .
LANCET, 1986, 1 (8476) :307-310
[2]  
BROOKS SM, 1972, J LAB CLIN MED, V79, P267
[3]  
BUTTNER J, 1976, CLIN CHEM, V22, P532
[4]   CONFIDENCE-INTERVALS RATHER THAN P-VALUES - ESTIMATION RATHER THAN HYPOTHESIS-TESTING [J].
GARDNER, MJ ;
ALTMAN, DG .
BMJ-BRITISH MEDICAL JOURNAL, 1986, 292 (6522) :746-750
[5]   OPTICAL FLUORESCENCE AND ITS APPLICATION TO AN INTRAVASCULAR BLOOD-GAS MONITORING-SYSTEM [J].
GEHRICH, JL ;
LUBBERS, DW ;
OPITZ, N ;
HANSMANN, DR ;
MILLER, WW ;
TUSA, JK ;
YAFUSO, M .
IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 1986, 33 (02) :117-132
[6]  
HANSEN JE, 1989, AM REV RESPIR DIS, V140, P402
[7]   VARIABILITY OF BLOOD-GASES, PULSE OXIMETER SATURATION, AND END-TIDAL CARBON-DIOXIDE PRESSURE IN STABLE, MECHANICALLY VENTILATED TRAUMA PATIENTS [J].
HESS, D ;
AGARWAL, NN .
JOURNAL OF CLINICAL MONITORING, 1992, 8 (02) :111-115
[8]   PROGRESS IN THE DEVELOPMENT OF A FLUORESCENT INTRAVASCULAR BLOOD-GAS SYSTEM IN MAN [J].
MAHUTTE, CK ;
SASSOON, CSH ;
MURO, JR ;
HANSMANN, DR ;
MAXWELL, TP ;
MILLER, WW ;
YAFUSO, M .
JOURNAL OF CLINICAL MONITORING, 1990, 6 (02) :147-157
[9]   DEVELOPMENT OF A PATIENT-DEDICATED, ON-DEMAND, BLOOD-GAS MONITOR [J].
MAHUTTE, CK ;
HOLODY, M ;
MAXWELL, TP ;
CHEN, PA ;
SASSE, SA .
AMERICAN JOURNAL OF RESPIRATORY AND CRITICAL CARE MEDICINE, 1994, 149 (04) :852-859
[10]  
MAHUTTE CK, 1991, CONT MANAGEMENT CRIT, V4, pCH2