A NEW ALGORITHM FOR MINIMIZING PACEMAKER POLARIZATION ARTIFACT - UNIVERSALLY APPLICABLE IN PERMANENT PACING SYSTEMS

被引:7
作者
CURTIS, AB
VANCE, F
QUIST, SM
DOMIJAN, A
KEIM, SG
DURAN, A
MILLER, K
机构
[1] VET AFFAIRS MED CTR,GAINESVILLE,FL
[2] TELECTRON PACING SYST,ENGLEWOOD,CO
来源
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY | 1991年 / 14卷 / 11期
关键词
PACEMAKERS; POLARIZATION ARTIFACT; PACEMAKER STIMULUS;
D O I
10.1111/j.1540-8159.1991.tb02770.x
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Polarization artifacts that result from pacing may interfere with analysis of paced evoked responses during, e.g., automatic threshold tracking. We have developed a method for reduction of such artifacts that relies on the introduction of pacing stimuli during the refractory period of unipolar or bipolar paced captured beats after previous identification of a refractory period "template" or baseline. The refractory pacing stimuli cannot capture the heart, and thus any deviation from the template is due to polarization artifact alone. The artifact amplitude is measured and the precharge duration of the triphasic stimulus waveform is changed each time until artifact is minimized, as detected by repeated reversals in the polarity of the polarization artifact. In a series of 11 patients with unipolar and bipolar permanent pacing leads, mean artifact before balancing was 1.44 +/- 0.84 mV, which was reduced to 0.44 +/- 0.30 mV after balancing (P = 0.001). Initial precharge duration was 3.2 msec by design; mean final precharge duration was 3.30 +/- 0.34 msec. This algorithm is universally applicable in permanent pacing systems, as it is valid in unipolar and bipolar pacing and it does not require an intrinsic cardiac rhythm.
引用
收藏
页码:1803 / 1808
页数:6
相关论文
共 8 条
[1]   AUTODIAGNOSTIC PACEMAKER [J].
AUERBACH, AA ;
FURMAN, S .
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1979, 2 (01) :58-68
[2]   MORPHOLOGY OF ENDOCARDIAL T-WAVES OF FUSION BEATS [J].
BOUTE, W ;
CALS, GLM ;
DENHEIJER, P ;
WITTKAMPF, FHM .
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1988, 11 (11) :1693-1697
[3]   THE VENTRICULAR DEPOLARIZATION GRADIENT - EFFECTS OF EXERCISE, PACING RATE, EPINEPHRINE, AND INTRINSIC HEART-RATE CONTROL ON THE RIGHT VENTRICULAR EVOKED-RESPONSE [J].
CALLAGHAN, F ;
VOLLMANN, W ;
LIVINGSTON, A ;
BOVEJA, B ;
ABELS, D .
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1989, 12 (07) :1115-1130
[4]   CHARACTERISTIC VARIATION IN EVOKED-POTENTIAL AMPLITUDE WITH CHANGES IN PACING STIMULUS STRENGTH [J].
CURTIS, AB ;
VANCE, F ;
MILLERSHIFRIN, K .
AMERICAN JOURNAL OF CARDIOLOGY, 1990, 66 (04) :416-422
[5]   AUTOMATIC REDUCTION OF STIMULUS POLARIZATION ARTIFACT FOR ACCURATE EVALUATION OF VENTRICULAR EVOKED-RESPONSES [J].
CURTIS, AB ;
VANCE, F ;
MILLER, K .
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1991, 14 (04) :529-537
[6]  
MUGICA J, 1973, CARDIAC PACING, P150
[7]  
PRESTON TA, 1973, CARDIAC PACING, P295
[8]  
STEINHAUS BM, 1990, 12TH P IEEE EMBS ANN, V12, P607