PERIPHERAL MR ANGIOGRAPHY WITH VARIABLE VELOCITY ENCODING - WORK IN PROGRESS

被引:30
作者
SWAN, JS [1 ]
WEBER, DM [1 ]
GRIST, TM [1 ]
WOJTOWYCZ, MM [1 ]
KOROSEC, FR [1 ]
MISTRETTA, CA [1 ]
机构
[1] UNIV WISCONSIN,CTR CLIN SCI,DEPT MED PHYS,MADISON,WI 53792
关键词
ARTERIES; MR; PERIPHERAL; BLOOD VESSELS; MAGNETIC RESONANCE (MR); IMAGE PROCESSING; PULSE SEQUENCES; TECHNOLOGY;
D O I
10.1148/radiology.184.3.1509072
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
An electrocardiographically triggered two-dimensional phase-contrast (PC) magnetic resonance angiographic pulse sequence was developed in which velocity encoding (VENC) was varied throughout an acquisition in response to changes in blood velocity during the cardiac cycle. This was done to better capture signal in the peripheral vasculature, where pulsatile flow degrades images. After reconstruction, a matched filter addition technique was applied to the cardiac phase images to obtain a single high-quality static image. Images were obtained of six healthy volunteers-with and without varying VENC-and contrast-to-noise ratio (C/N) calculations were performed for the added images. Varying VENC significantly improved vascular signal from small and large vessels (P < .02), but it was most helpful for small vessels, for which the C/N increased by as much as 260% (average increase, 149%). These preliminary findings suggest that variable VENC can enhance the signal from the small and large peripheral blood vessels in cardiac-gated PC acquisitions.
引用
收藏
页码:813 / 817
页数:5
相关论文
共 7 条
  • [1] BUONOCORE MH, 1991, BOOK ABSTRACTS SOC M, P804
  • [2] MAGNETIC-RESONANCE ANGIOGRAPHY
    DUMOULIN, CL
    HART, HR
    [J]. RADIOLOGY, 1986, 161 (03) : 717 - 720
  • [3] ECG-OPTIMIZED PHASE-CONTRAST LINE-SCANNED MR ANGIOGRAPHY
    KOROSEC, FR
    MISTRETTA, CA
    TURSKI, PA
    [J]. MAGNETIC RESONANCE IN MEDICINE, 1992, 24 (02) : 221 - 235
  • [4] BLOOD-FLOW IMAGING BY CINE MAGNETIC-RESONANCE
    NAYLER, GL
    FIRMIN, DN
    LONGMORE, DB
    [J]. JOURNAL OF COMPUTER ASSISTED TOMOGRAPHY, 1986, 10 (05) : 715 - 722
  • [5] NEWMEYER MM, 1988, CLIN APPLICATIONS DO, P330
  • [6] PELC NJ, 1989, BOOK ABSTRACTS SOC M, P101
  • [7] SCHWARTZ M, 1975, SIGNAL PROCESSING DI, P251