CRYOGENICALLY COOLED ADSORBENT TRAP FOR OFF-LINE SUPERCRITICAL FLUID EXTRACTION

被引:45
作者
ASHRAFKHORASSANI, M [1 ]
HOUCK, RK [1 ]
LEVY, JM [1 ]
机构
[1] SUPREX CORP,125 WILLIAM PITT WAY,PITTSBURGH,PA 15238
关键词
D O I
10.1093/chromsci/30.9.361
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
The off-line collection of extracted analytes after supercritical fluid extraction (SFE) is an important step for successful extraction. Percent recoveries of various analytes using three different methods of off-line collection are reported. Poor collection efficiency is shown when the sample was collected in an empty vial. High collection efficiency (b 90%) of standards in a vial with the solvent is obtained only when a low flow rate of liquid CO2(` 1 mL/min) is passed through the extraction vessel. Collection of the same analytes in a cryogenically cooled vessel filled with adsorbent shows an efficiency of greater than 90%. The flow rate of liquid CO2through an extraction vessel does not have any effect on the collection efficiency. Collection of volatile analytes in a cryogenically cooled adsorbent trap requires low temperature trapping (` -40 C). Also, the type of liquid solvent used to wash the extracted analytes in a cryogenically cooled adsorbent trap is of importance. © Oxford University Press 1992.