VERIFICATION OF GAMMA-SPECTROSCOPY PROGRAMS - ACCURACY AND DETECTABILITY

被引:7
作者
MERCIER, MT [1 ]
KOSKELO, MJ [1 ]
机构
[1] CANBERRA NUCL PROD GRP,MERIDEN,CT 06450
来源
JOURNAL OF RADIOANALYTICAL AND NUCLEAR CHEMISTRY-ARTICLES | 1992年 / 160卷 / 01期
关键词
D O I
10.1007/BF02041673
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
The newly revised ANSI N42.14(1,2) has provided analysis software developers with a set of well defined, consistent and unbiased procedures designed to evaluate the accuracy and limitations of peak search and peak area analysis programs. This work uses two of the procedures outlined in this standard to evaluate five peak analysis algorithms currently in use in Canberra and Nuclear Data software packages. The first procedure examines a program's behavior as the centroid separation and peak height ratio of a doublet are varied. A previous review of these data3 demonstrated significant peak area inaccuracies at peak separations at or below 1.5 FWHM. We will discuss improvements made to some of these programs and the impact on the doublet results. The second procedure examines a program's behavior as the Compton continuum beneath a fixed peak area is increased. For the same five algorithms we will discuss the dependence of peak area on Compton continuum and also explore the limits of peak detectability.
引用
收藏
页码:233 / 243
页数:11
相关论文
共 3 条
[2]  
GEHRKE RJ, 1990, TESTS MEASURE PERFOR, P19
[3]   VERIFICATION OF GAMMA-SPECTROSCOPY PROGRAMS - A STANDARDIZED APPROACH [J].
KOSKELO, MJ ;
MERCIER, MT .
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 1990, 299 (1-3) :318-321