VOLTAMMETRIC QUANTIFICATION AND SPECIATION OF MERCURY-COMPOUNDS

被引:50
作者
AGRAZ, R [1 ]
SEVILLA, MT [1 ]
HERNANDEZ, L [1 ]
机构
[1] UNIV AUTONOMA MADRID,DEPT ANALYT CHEM & INSTRUMENTAL ANAL,E-28049 MADRID,SPAIN
来源
JOURNAL OF ELECTROANALYTICAL CHEMISTRY | 1995年 / 390卷 / 1-2期
关键词
MERCURY ASSAY; ORGANOMETALLIC COMPOUNDS; MODIFIED CARBON ELECTRODES; CYCLIC VOLTAMETRY;
D O I
10.1016/0022-0728(95)03955-G
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
The use of a carbon paste electrode modified with a thiolic resin for the determination of inorganic mercury and organomercury compounds, present simultaneously in a sample, is described. The compounds are first preconcentrated at the electrode surface by means of a purely chemical reaction with the modifier on the electrode surface. The high affinity of the modifier for the mercury compounds ensures low limits of detection and determination. Differentiation between several mercury species is possible by control of the reduction potential applied to the working electrode. This selective reduction results in the formation of atomic mercury at the electrode surface which can be determined with a very high sensitivity by means of its re-oxidation wave in cyclic voltammetry. Optimization of the instrumental parameters and evidence for the reduction processes are discussed. Analysis of inorganic mercury in the presence of methylmercury, with a detection limit of 4 mu g Hg l(-1), and of methylmercury in the presence of inorganic mercury, with a detection limit of 2 mu g Hg l(-1), is described in detail. In both cases the preconcentration time is 6 min. Other organomercury species can also be quantified. Application of the method to environmental aquatic samples is discussed.
引用
收藏
页码:47 / 57
页数:11
相关论文
共 55 条
[41]   ASSAY OF PHENYLMERCURY SALTS IN PHARMACEUTICAL PRODUCTS BY HIGH-PERFORMANCE LIQUID-CHROMATOGRAPHY OF THE MORPHOLINEDITHIOCARBAMATE DERIVATIVE [J].
PARKIN, JE .
JOURNAL OF CHROMATOGRAPHY, 1987, 407 :389-392
[42]   HIGH-PERFORMANCE LIQUID-CHROMATOGRAPHY OF MERCURY AND PHENYLMERCURY AS THE N-DISUBSTITUTED DITHIOCARBAMATE COMPLEXES [J].
PARKIN, JE .
JOURNAL OF CHROMATOGRAPHY, 1989, 472 (02) :401-405
[43]   ANODIC STRIPPING VOLTAMMETRY OF MERCURY(2) AT GRAPHITE ELECTRODE [J].
PERONE, SP ;
KRETLOW, WJ .
ANALYTICAL CHEMISTRY, 1965, 37 (8P1) :968-&
[44]  
PHILLIPS RJ, 1987, ANAL CHEM, V50, P1504
[45]  
RIEVAJ M, 1992, ANALYST, V117, P117
[46]   SPECIATION OF MERCURY-COMPOUNDS BY TLC AND HIGH-PERFORMANCE LIQUIDCHROMATOGRAPHY ATOMIC ABSORPTION-SPECTROSCOPY [J].
ROBINSON, JW ;
BOOTHE, ED .
SPECTROSCOPY LETTERS, 1984, 17 (11) :673-688
[47]  
SANE KV, 1991, INDIAN J CHEM, V29, P504
[48]   ACCUMULATION VOLTAMMETRY OF COPPER(II) USING A CARBON PASTE ELECTRODE MODIFIED WITH DI-8-QUINOLYL DISULFIDE [J].
SUGAWARA, K ;
TANAKA, S ;
TAGA, M .
ANALYST, 1991, 116 (02) :131-134
[49]  
SUZUKI Y, 1985, N PACIFIC TOTAL ORGA
[50]   SELECTIVE DETERMINATION OF CR(VI) OXYANIONS USING A POLY-3-METHYLTHIOPHENE-MODIFIED ELECTRODE [J].
TEASDALE, PR ;
SPENCER, MJ ;
WALLACE, GG .
ELECTROANALYSIS, 1989, 1 (06) :541-547