APPLICATION OF ADSORPTIVE STRIPPING VOLTAMMETRY (ADSV) FOR THE ANALYSIS OF TRACE-METALS IN BRINE .2. DEVELOPMENT AND EVALUATION OF A FLOW-INJECTION SYSTEM

被引:17
作者
ROMANUS, A
MULLER, H
KIRSCH, D
机构
[1] TH CARL SCHORLEMMER LEUNA MERSEBURG, FACHBEREICH CHEM, INST ANALYT & UMWELTCHEM, OTTO NUSCHKE STR, O-4200 MERSEBURG, GERMANY
[2] BUNA AG, ZENT BEREICH FORSCH & ENTWICKLUNG, HAUPTABT ANALYT, W-4212 SCHKOPAU, GERMANY
来源
FRESENIUS JOURNAL OF ANALYTICAL CHEMISTRY | 1991年 / 340卷 / 06期
关键词
D O I
10.1007/BF00321584
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
A flow system for quantifying traces of heavy metals in brine by adsorptive stripping voltammetry (AdSV) is described. The system is developed and optimized for simplified, rapid sample processing using either linear scan or differential pulse waveforms. The advantages of the system include improved efficiency of sample preconcentration during the accumulation step, minimal sample pretreatment, simplified operation and sample deoxygenation, possible medium exchange between the accumulation and stripping steps, ready adaptility to automation and on-line monitoring. Minimum detection limits are less than 5-mu-g/l for all trace metals with precisions better than 7%.
引用
收藏
页码:371 / 376
页数:6
相关论文
共 18 条
[1]  
BOLTZ HJ, 1987, ELEKTROPRAKTIKER, V41, P310
[2]   GAS-DIFFUSION SEPARATION AND FLOW-INJECTION POTENTIOMETRY - A FRUITFUL ALLIANCE OF ANALYTICAL METHODS [J].
FRENZEL, W .
FRESENIUS JOURNAL OF ANALYTICAL CHEMISTRY, 1990, 336 (01) :21-28
[3]   DETERMINATION OF URANIUM(VI) IN SEAWATER BY MEANS OF AUTOMATED FLOW CONSTANT-CURRENT CATHODIC STRIPPING AT CARBON-FIBER ELECTRODES [J].
HUA, C ;
JAGNER, D ;
RENMAN, L .
ANALYTICA CHIMICA ACTA, 1987, 197 :265-273
[4]   ADSORPTIVE STRIPPING VOLTAMMETRY IN TRACE ANALYSIS [J].
KALVODA, R ;
KOPANICA, M .
PURE AND APPLIED CHEMISTRY, 1989, 61 (01) :97-112
[5]   LIMIT OF DECISION, LIMIT OF DETECTION AND LIMIT OF DETERMINATION OF ANALYTICAL METHODS [J].
LUTHARDT, M ;
THAN, E ;
HECKENDORFF, H .
FRESENIUS ZEITSCHRIFT FUR ANALYTISCHE CHEMIE, 1987, 326 (04) :331-339
[6]   AN ASSESSMENT OF THE RATE OF OCCURRENCE OF HIGH HYDROGEN CONTENT IN A CHLORINE PRODUCTION PLANT [J].
NIELSEN, DS ;
PLATZ, O .
RELIABILITY ENGINEERING & SYSTEM SAFETY, 1983, 4 (01) :1-18
[7]   DEOXYGENATION OF SUPPORTING ELECTROLYTES IN STRIPPING VOLTAMMETRY BY GLUCOSE AND CO-IMMOBILIZED GLUCOSE-OXIDASE AND CATALASE IN A FLOW SYSTEM [J].
RISINGER, L ;
YANG, XR ;
JOHANSSON, G .
ANALYTICA CHIMICA ACTA, 1987, 200 (01) :313-318
[8]  
ROMANUS A, 1991, IN PRESS FRESENIUS J
[9]   EFFECT OF VARIOUS MEASURING TECHNIQUES ON THE RESPONSE OF A POLAROGRAPHIC HIGH-PERFORMANCE LIQUID-CHROMATOGRAPHIC DETECTOR [J].
STULIK, K ;
PACAKOVA, V ;
PODOLAK, M .
JOURNAL OF CHROMATOGRAPHY, 1983, 262 (JUN) :85-94
[10]   THE CONTINUOUS REMOVAL OF OXYGEN FROM FLOWING SOLUTIONS [J].
TROJANEK, A ;
HOLUB, K .
ANALYTICA CHIMICA ACTA, 1980, 121 (DEC) :23-28