SENSITIVITY AND SELECTIVITY OF INTRANEURAL STIMULATION USING A SILICON ELECTRODE ARRAY

被引:72
作者
RUTTEN, WLC [1 ]
VANWIER, HJ [1 ]
PUT, JHM [1 ]
机构
[1] TWENTE UNIV TECHNOL, DEPT ELECT ENGN, DIV BIOMED ENGN, 7500 AE ENSCHEDE, NETHERLANDS
关键词
D O I
10.1109/10.76386
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Artificial electrical stimulation of peripheral nerves needs the development of multielectrode devices which stimulate individual fibers or small groups in a selective and sensitive way. To this end, a multielectrode array in silicon technology has been developed, as well as experimental paradigms and model calculations for sensitivity and selectivity measures. The array consists of twelve platinum electrode sites (10 x 50-mu-m at 50-mu-m interdistance) on a 45-mu-m thick tip-shaped silicon substrate and a Si3N4 insulating glass cover layer. The tip is inserted in the peroneal nerve of the rat during acute experiments to stimulate alpha motor fibers of the extensor digitorum longus muscle. Sensitivity calculations and experiments show a cubic dependence of the number of stimulated motor units on current amplitude of the stimulatory pulse (recruitment curves), starting at single motor level. Selectivity was tested by a method based on the refractory properties of neurons. At the lowest stimulus levels (for one motor unit) selectivity is maximal when two electrodes are separated by 200-250-mu-m, which was estimated also on theoretical grounds. The study provides clues for future designs of two- and three-dimensional devices.
引用
收藏
页码:192 / 198
页数:7
相关论文
共 24 条
[1]   ACUTE AND CHRONIC IMPLANTATION OF COILED WIRE INTRANEURAL ELECTRODES DURING CYCLICAL ELECTRICAL-STIMULATION [J].
BOWMAN, BR ;
ERICKSON, RC .
ANNALS OF BIOMEDICAL ENGINEERING, 1985, 13 (01) :75-93
[2]  
CLARK LD, 1987, 9TH P IEEE BME C BOS, P1660
[3]   PROPERTIES OF MOTOR UNITS IN FAST AND SLOW SKELETAL MUSCLES OF RAT [J].
CLOSE, R .
JOURNAL OF PHYSIOLOGY-LONDON, 1967, 193 (01) :45-&
[4]   LONG-TERM NEUROELECTRIC SIGNAL RECORDING FROM SEVERED NERVES [J].
DE LUCA, CJ ;
GILMORE, LD ;
BLOOM, LJ ;
THOMSON, SJ ;
CUDWORTH, AL ;
GLIMCHER, MJ .
IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 1982, 29 (06) :393-403
[5]  
DONALDSON ND, 1986, MED BIOL ENG COMPUT, V24, P41, DOI 10.1007/BF02441604
[6]   A PERIPHERAL-NERVE INFORMATION TRANSDUCER FOR AMPUTEES - LONG-TERM MULTICHANNEL RECORDINGS FROM RABBIT PERIPHERAL-NERVES [J].
EDELL, DJ .
IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 1986, 33 (02) :203-214
[7]   SPECIFIC RESISTANCE OF BIOLOGICAL MATERIAL-A COMPENDUM OF DATA FOR BIOMEDICAL ENGINEER AND PHYSIOLOGIST [J].
GEDDES, LA ;
BAKER, LE .
MEDICAL & BIOLOGICAL ENGINEERING, 1967, 5 (03) :271-&
[8]   ORGANIZATION OF NEURONS FORMING THE FEMORAL, SCIATIC, COMMON PERONEAL AND TIBIAL NERVES IN RATS AND MONKEYS [J].
JANJUA, MZ ;
LEONG, SK .
BRAIN RESEARCH, 1984, 310 (02) :311-323
[9]   DISTINGUISHABILITY OF FUNCTIONALLY DISTINCT EVOKED NEUROELECTRIC SIGNALS ON THE SURFACE OF A NERVE [J].
LICHTENBERG, BK ;
DE LUCA, CJ .
IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 1979, 26 (04) :228-237
[10]   ANALYSIS OF A MODEL FOR EXCITATION OF MYELINATED NERVE [J].
MCNEAL, DR .
IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 1976, 23 (04) :329-337