MAXIMIZING SETUP ACCURACY USING PORTAL IMAGES AS APPLIED TO A CONFORMAL BOOST TECHNIQUE FOR PROSTATIC-CANCER

被引:149
作者
BIJHOLD, J [1 ]
LEBESQUE, JV [1 ]
HART, AAM [1 ]
VIJLBRIEF, RE [1 ]
机构
[1] NETHERLANDS CANC INST,ANTONI VANLEEUWENHOEK HUIS,DEPT RADIOTHERAPY,PLESMANLAAN 121,1066 CX AMSTERDAM,NETHERLANDS
关键词
RADIATION THERAPY; RADIATION FIELD ALIGNMENT; PATIENT SETUP VERIFICATION; DECISION RULES; QUALITY ASSURANCE; PORTAL IMAGING; CONFORMAL FIELD THERAPY; PROSTATIC CANCER;
D O I
10.1016/0167-8140(92)90233-K
中图分类号
R73 [肿瘤学];
学科分类号
100214 ;
摘要
A design procedure of a patient setup verification protocol based upon frequent digital acquisition of portal images is demonstrated with an application for conformal prostatic boost fields. The protocol aims at the elimination of large systematic deviations in the patient setup and includes decision rules which indicate when correction of the patient setup is needed. The decision rules were derived from the results of a theoretical and quantitative analysis of patient setup variations measured in three pelvic fields (one anterior-posterior and two lateral fields) of 105 fractions for nine patients. Deviations in the patient positioning, derived from one field, were quantified as two-dimensional (2-D) displacement vectors in the plane perpendicular to the beam axis by alignment of anatomical features in the portal and the simulator image. The magnitude of the overall setup variations along the anterior-posterior, superior-inferior and lateral directions varied between 2.6 and 3 mm (1 S.D.). Inter- and intra-treatment variations could be separated, both having equal magnitudes of 1.7 to 2.2 mm (1 S.D.). In addition, intra-treatment variations appeared to be predictable which was a prerequisite for the development of the decision rules. The 2-D setup deviations, measured in the three fields of one fraction were strongly correlated and a 3-D displacement vector was calculated. Utilization of this 3-D vector in a setup verification protocol may lead to an early detection of systematic setup deviations.
引用
收藏
页码:261 / 271
页数:11
相关论文
共 19 条
[1]   RADIATION-FIELD EDGE-DETECTION IN PORTAL IMAGES [J].
BIJHOLD, J ;
GILHUIJS, KGA ;
VANHERK, M ;
MEERTENS, H .
PHYSICS IN MEDICINE AND BIOLOGY, 1991, 36 (12) :1705-1710
[2]   FAST EVALUATION OF PATIENT SET-UP DURING RADIOTHERAPY BY ALIGNING FEATURES IN PORTAL AND SIMULATOR IMAGES [J].
BIJHOLD, J ;
VANHERK, M ;
VIJLBRIEF, R ;
LEBESQUE, JV .
PHYSICS IN MEDICINE AND BIOLOGY, 1991, 36 (12) :1665-1679
[3]   QUALITY-CONTROL OF RADIOTHERAPY DURING PELVIC IRRADIATION [J].
GRIFFITHS, SE ;
KHOURY, GG ;
EDDY, A .
RADIOTHERAPY AND ONCOLOGY, 1991, 20 (03) :203-206
[4]  
HEUKELOM S, 1992, UNPUB RADIOTHER ONCO
[5]   ACCURACY IN RADIATION-FIELD ALIGNMENT IN HEAD AND NECK-CANCER - A PROSPECTIVE-STUDY [J].
HUIZENGA, H ;
LEVENDAG, PC ;
DEPORRE, PMZR ;
VISSER, AG .
RADIOTHERAPY AND ONCOLOGY, 1988, 11 (02) :181-187
[6]   A NEW SYSTEM FOR PATIENT FIXATION IN RADIOTHERAPY [J].
JAKOBSEN, A ;
IVERSEN, P ;
GADEBERG, C ;
HANSEN, JL ;
HJELMHANSEN, M .
RADIOTHERAPY AND ONCOLOGY, 1987, 8 (02) :145-151
[7]  
KENDALL MG, 1968, DESIGN ANAL TIME SER, V3, P265
[8]   THE SIMULTANEOUS BOOST TECHNIQUE - THE CONCEPT OF RELATIVE NORMALIZED TOTAL DOSE [J].
LEBESQUE, JV ;
KEUS, RB .
RADIOTHERAPY AND ONCOLOGY, 1991, 22 (01) :45-55
[9]   DETECTION OF SYSTEMATIC PATIENT SETUP ERRORS BY PORTAL FILM ANALYSIS [J].
LEBESQUE, JV ;
BEL, A ;
BIJHOLD, J ;
HART, AAM .
RADIOTHERAPY AND ONCOLOGY, 1992, 23 (03) :198-198
[10]   BEAMS EYE VIEW BASED PROSTATE TREATMENT PLANNING - IS IT USEFUL [J].
LOW, NN ;
VIJAYAKUMAR, S ;
ROSENBERG, I ;
RUBIN, S ;
VIRUDACHALAM, R ;
SPELBRING, DR ;
CHEN, GTY .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 1990, 19 (03) :759-768