QUANTITATIVE-ANALYSIS OF ELECTRONIC-GRADE ANHYDROUS HYDROGEN-CHLORIDE BY SEALED INDUCTIVELY-COUPLED PLASMA-ATOMIC EMISSION-SPECTROSCOPY

被引:14
作者
JACKSIER, T [1 ]
BARNES, RM [1 ]
机构
[1] UNIV MASSACHUSETTS,LEDERLE GRAD RES CTR,DEPT CHEM,AMHERST,MA 01003
关键词
SEALED DISCHARGE; GAS ANALYSIS; INDUCTIVELY COUPLED PLASMA; HYDROGEN CHLORIDE; EMISSION SPECTROSCOPY;
D O I
10.1039/ja9940901299
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
Anhydrous hydrogen chloride at concentrations of up to 100% has been introduced into a sealed inductively coupled. plasma system for quantitative spectrochemical analysis. Vapour-phase sampling of monobutyltin trichloride was developed to calibrate tin and carbon. The addition of chlorine as a modifier gas was required in a ratio of 1:1 to maintain stable and reproducible emission signals. Under flowing conditions detection limits for tin and carbon were 49 and 271 (ng g(-1)), respectively, in a 16% v/v HCl-Cl-2 argon plasma. Impurities identified qualitatively included Al, C, Ca, Cr, Fe, Ni and Sn. In addition, the bulk temperature of the plasma was determined to be 10500 +/- 1500 K.