ADIABATIC SLICE-SELECTIVE RF PULSES AND A SINGLE-SHOT ADIABATIC LOCALIZATION PULSE SEQUENCE

被引:28
作者
SLOTBOOM, J [1 ]
BOVEE, WMMJ [1 ]
机构
[1] DELFT UNIV TECHNOL,DEPT APPL PHYS,2600 GA DELFT,NETHERLANDS
来源
CONCEPTS IN MAGNETIC RESONANCE | 1995年 / 7卷 / 03期
关键词
D O I
10.1002/cmr.1820070303
中图分类号
O64 [物理化学(理论化学)、化学物理学];
学科分类号
070304 ; 081704 ;
摘要
The localization quality of frequency-selective adiabatic radio frequency (rf) pulses is substantially less affected by rf inhomogeneity than is the localization of amplitude-modulated pulses. For this reason, adiabatic pulses are popular for use in experiments carried out with rf coils that produce inhomogeneous rf fields. Starting with the Schrodinger equation, the Bloch equations are derived, neglecting relaxation. The adiabatic principle is introduced, and it is demonstrated that accurate adiabatic localization is possible with simple schemes for amplitude and frequency modulation. Adiabatic pulses perform well only in an rf amplitude range limited by a lower and an upper bound. Phase-compensated, slice-selective adiabatic 2 pi pulses are used in a variant of the single-shot localization method, SADLOVE, for which a theoretical quantitative model of the localization performance is derived. A simple variant of the VERSE method to reduce the rf pulse energy of adiabatic pulses is presented. (C) 1995 John Wiley & Sons, Inc.
引用
收藏
页码:193 / 217
页数:25
相关论文
共 23 条
[11]  
CONOLLY S, 1989, 8TH ANN M SOC MAGN R, P861
[12]   LOCALIZED PROTON SPECTROSCOPY USING STIMULATED ECHOES [J].
FRAHM, J ;
MERBOLDT, KD ;
HANICKE, W .
JOURNAL OF MAGNETIC RESONANCE, 1987, 72 (03) :502-508
[13]   A METHOD FOR CHARACTERIZING LOCALIZATION TECHNIQUES IN VOLUME SELECTED NUCLEAR-MAGNETIC-RESONANCE SPECTROSCOPY [J].
KEEVIL, SF ;
PORTER, DA ;
SMITH, MA .
PHYSICS IN MEDICINE AND BIOLOGY, 1990, 35 (07) :821-834
[14]  
LANDAU LD, 1979, LEHRBUCH THEORETISCH, V3
[15]  
Ordidge R, 1985, MAGNETIC RESONANCE B, P387
[16]   HIGHLY SELECTIVE PI/2 AND PI-PULSE GENERATION [J].
SILVER, MS ;
JOSEPH, RI ;
HOULT, DI .
JOURNAL OF MAGNETIC RESONANCE, 1984, 59 (02) :347-351
[17]   SPATIALLY SELECTIVE RF PULSES AND THE DIGITIZATION ON THEIR PERFORMANCE [J].
SLOTBOOM, J ;
CREYGHTON, JHN ;
KORBEE, D ;
MEHLKOPF, AF ;
BOVEE, WMMJ .
MAGNETIC RESONANCE IN MEDICINE, 1993, 30 (06) :732-740
[18]   A SINGLE-SHOT LOCALIZATION PULSE SEQUENCE SUITED FOR COILS WITH INHOMOGENEOUS RF FIELDS USING ADIABATIC SLICE-SELECTIVE RF PULSES [J].
SLOTBOOM, J ;
MEHLKOPF, AF ;
BOVEE, WMMJ .
JOURNAL OF MAGNETIC RESONANCE, 1991, 95 (02) :396-404
[19]   PROTON-RESONANCE SPECTROSCOPY STUDY OF THE EFFECTS OF L-ORNITHINE-L-ASPARTATE ON THE DEVELOPMENT OF ENCEPHALOPATHY, USING LOCALIZATION PULSES WITH REDUCED SPECIFIC ABSORPTION RATE [J].
SLOTBOOM, J ;
VOGELS, BAPM ;
DEHAAN, JG ;
CREYGHTON, JHN ;
QUACK, G ;
CHAMULEAU, RAFM ;
BOVEE, WMMJ .
JOURNAL OF MAGNETIC RESONANCE SERIES B, 1994, 105 (02) :147-156
[20]  
SLOTBOOM J, 1991, 8TH P ANN C EUR SOC, P273