The analytical behaviour of dichloromethane, pentane, and hexane solutions was studied through their evaporation in a micro-Kuderna-Danish concentrator. Initial sample concentration, final sample volume, matrix composition, system geometry and temperature gradients influence the process greatly. In general, solute loss increases exponentially when the final volume decreases. During the concentration process with very dilute solutions, the loss of solutes can be higher than 50% if the final volume reached is lower than 200 mu l. This means a decrease in the quantification of the process, even though it is seemingly kept under control (low R.S.D.). The solute losses are not just due to the decrease in the number of plates during the reflux disappearance, but also to the probability that droplets of the evaporating sample are projected against clean and dry walls of low retention capacity. This hypothesis may explain the observed effects when concentrating more concentrated samples, or samples with a high content of some solutes. A more advantageous design for the micro-Kuderna-Danish receptor flask is also proposed. The replacement of the cylindrical flask by a flask with a conical bottom permits a decrease in the volume of losses and an improvement in the quantification of the process.
引用
收藏
页码:41 / 55
页数:15
相关论文
共 22 条
[1]  
AHRENSTLARSEN B, 1964, WALLERSTEIN LAB COMM, V27, P41
[2]   CONVENIENT METHOD FOR MULTIPLE EXTRACTION OF VOLATILE FLAVOR COMPONENTS FROM FOOD SLURRIES AND PULPS USING A 2-CHAMBERED GLASS BOMB EXTRACTOR AND DICHLORODIFLUOROMETHANE (FREON-12) SOLVENT [J].
BLAKESLEY, CN ;
LOOTS, J .
JOURNAL OF AGRICULTURAL AND FOOD CHEMISTRY, 1977, 25 (04) :961-963
[3]   NEW SENSITIVE METHOD FOR THE EXAMINATION OF THE VOLATILE FLAVOR FRACTION OF CABERNET-SAUVIGNON WINES [J].
BOISON, JOK ;
TOMLINSON, RH .
JOURNAL OF CHROMATOGRAPHY, 1990, 522 :315-327
[4]  
DUNGES D, 1979, PRACHROMATOGRAPHISCH
[5]   MATRIX EFFECTS AND SOLUTE DISCRIMINATION WHEN INJECTING DIRTY SAMPLES IN CAPILLARY COLUMNS - COMPARATIVE-STUDY BETWEEN CLASSICAL SPLIT AND SPLITLESS INJECTIONS [J].
FERREIRA, V ;
ESCUDERO, A ;
SALAFRANCA, J ;
FERNANDEZ, P ;
CACHO, J .
JOURNAL OF CHROMATOGRAPHY A, 1993, 655 (02) :257-266
[6]   EFFECT OF DIRT ON QUANTITATIVE-ANALYSES BY CAPILLARY GAS-CHROMATOGRAPHY WITH SPLITLESS INJECTION [J].
GROB, K ;
BOSSARD, M .
JOURNAL OF CHROMATOGRAPHY, 1984, 294 (JUN) :65-75
[7]  
GROB K, 1992, HRC-J HIGH RES CHROM, V15, P190
[8]  
GROB K, 1992, HRC-J HIGH RES CHROM, V15, P399
[9]   FILM THICKNESS REQUIRED OF THE STATIONARY PHASE IN THE SEPARATION COLUMN WHEN USING LARGE RETENTION GAPS IN CAPILLARY GAS-CHROMATOGRAPHY [J].
GROB, K .
JOURNAL OF CHROMATOGRAPHY, 1985, 328 (JUN) :55-61
[10]   SAMPLE RECONCENTRATION BY COLUMN-EXTERNAL SOLVENT EVAPORATION OR INJECTION OF LARGE VOLUMES INTO GAS-CHROMATOGRAPHIC CAPILLARY COLUMNS [J].
GROB, K ;
MULLER, E .
JOURNAL OF CHROMATOGRAPHY, 1987, 404 (02) :297-305