OPTIMIZATION OF DATA HANDLING AND DETECTION CONDITIONS FOR AUTOMATED CHROMATOGRAPHIC ASSAY SOFTWARE

被引:10
作者
CARDOT, PJP
TROLLIARD, P
GUERNETNIVAUD, E
机构
[1] Laboratoire de Chimie Analytique et d'Electrochimie Organiques, Université Paris-Sud, Centre d'Etudes Pharmaceutiques, Châtenay-Malabry, 92296
关键词
COLUMN LIQUID CHROMATOGRAPHY; AUTOMATED PEAK DETECTION; DISCRETE FOURIER TRANSFORM; FREQUENCY ACQUISITION AND FILTERING; PEAK DETECTION;
D O I
10.1007/BF02275919
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
The principle of automated chromatographic peak detection and analysis software is summarized, and critical steps are systematically studied. As the only parameter to be entered is the acquisition frequency, evaluation of its effect on software performance is discussed. In the case of relatively noisy chromatograms, it is shown experimentally that numerous points per peak have to be taken, leading to quite fast computer acquisition procedures. The use of discrete Fourier transform filtration techniques can modify peak shapes and a comparative study evaluates the relative errors induced in the shapes and characteristics of the chromatographic profiles. Optimisation of filtering conditions is achieved and it is shown that for a filter position only 2% of the Nyquist frequency no deformation occurs in the chromatographic profile. Detection of the start and finish of chromatographic peaks is optimized according to a simple four step iterative procedure. In the case of simulations, the difference between the values used to simulate peaks and those calculated by the software are less than 1%.
引用
收藏
页码:361 / 368
页数:8
相关论文
共 15 条
[1]   A FULLY AUTOMATED CHROMATOGRAPHIC PEAK DETECTION AND TREATMENT SOFTWARE FOR MULTIUSER MULTITASK COMPUTERS [J].
CARDOT, PJP ;
TROLLIARD, P ;
TEMBELY, S .
JOURNAL OF PHARMACEUTICAL AND BIOMEDICAL ANALYSIS, 1990, 8 (8-12) :755-759
[2]   EFFECT OF PEAK SENSING AND RANDOM NOISE ON PRECISION AND ACCURACY OF STATISTICAL MOMENT ANALYSES FROM DIGITAL CHROMATOGRAPHIC DATA [J].
CHESLER, SN ;
CRAM, SP .
ANALYTICAL CHEMISTRY, 1971, 43 (14) :1922-&
[3]   EFFECTS OF DIGITAL-FILTERS ON CHROMATOGRAPHIC SIGNALS [J].
CRAM, SP ;
CHESLER, SN ;
BROWN, AC .
JOURNAL OF CHROMATOGRAPHY, 1976, 126 (NOV3) :279-300
[5]  
DYSON N, 1990, RSC CHROMATOGRAPHY M
[6]  
ESCOFFIER JL, 1983, CHROMATOGRAPHIA, V15, P543
[7]   A REVIEW OF THE EXPONENTIALLY MODIFIED GAUSSIAN (EMG) FUNCTION - EVALUATION AND SUBSEQUENT CALCULATION OF UNIVERSAL DATA [J].
FOLEY, JP ;
DORSEY, JG .
JOURNAL OF CHROMATOGRAPHIC SCIENCE, 1984, 22 (01) :40-46
[8]   REVIEW OF THE EXPONENTIALLY MODIFIED GAUSSIAN (EMG) FUNCTION SINCE 1983 [J].
JEANSONNE, MS ;
FOLEY, JP .
JOURNAL OF CHROMATOGRAPHIC SCIENCE, 1991, 29 (06) :258-266
[9]   PRACTICAL CONSIDERATIONS FOR DIGITIZING ANALOG SIGNALS [J].
KELLY, PC ;
HORLICK, G .
ANALYTICAL CHEMISTRY, 1973, 45 (03) :518-527
[10]   RESULTS OF A COOPERATIVE STUDY COMPARING THE PRECISION OF PEAK HEIGHT AND AREA MEASUREMENTS IN LIQUID-CHROMATOGRAPHY [J].
MCCOY, RW ;
AIKEN, RL ;
PAULS, RE ;
ZIEGEL, ER ;
WOLF, T ;
FRITZ, GT ;
MARMION, DM .
JOURNAL OF CHROMATOGRAPHIC SCIENCE, 1984, 22 (10) :425-431