EFFECT OF SPECTRAL RESOLUTION, DETECTOR LINEARITY AND CHROMATOGRAPHIC RESOLUTION ON PEAK PURITY CALCULATIONS

被引:4
作者
ANDREWS, RW
RICHARDSON, H
机构
[1] Waters Chromatography Division, Millipore Corp., Milford, MA 01757
关键词
D O I
10.1016/S0021-9673(94)89096-X
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
The detection of inhomogeneity in chromatographic peaks is one of the principal benefits of photodiode array detection for HPLC. The peak homogeneity can be estimated by comparing the angles between the vector representations of the instantaneous spectrum and the peak apex spectrum, and the instantaneous spectrum and the spectral noise. When the spectral or purity angle exceeds the noise angle, the peak is not homogeneous because the differences between the instantaneous spectra and the apex spectrum cannot be explained by statistical variation of the apex spectrum. In this paper, we consider the impact of detector linearity and slit width on the purity (homogeneity) measurements of a separation of a series of compounds related to vanillin. The peak purity angle, i.e., the spectral contrast of spectra within a chromatographic peak, varies with the maximum absorbance. At small maximum absorbances (less than 0.1 AU), the variation in purity angle is dominated by system noise. At large maximum absorbances (greater than 0.5 AU), the variation in purity angle is dominated by the photometric uncertainty. The smallest measurable purity angle for a homogeneous peak is inversely proportional to the slit width and spectral bandpass of the polychrometer of the photodiode array detector.
引用
收藏
页码:3 / 8
页数:6
相关论文
共 17 条
[1]   RECENT DEVELOPMENTS IN MULTICHANNEL, PHOTODIODE-ARRAY, OPTICAL LC DETECTION [J].
ALFREDSON, T ;
SHEEHAN, T .
JOURNAL OF CHROMATOGRAPHIC SCIENCE, 1986, 24 (11) :473-482
[2]  
CAMERON G, 1993, CHEM AUST, V60, P288
[3]   STRATEGIES FOR PEAK-PURITY ASSESSMENT IN LIQUID-CHROMATOGRAPHY [J].
CASTLEDINE, JB ;
FELL, AF .
JOURNAL OF PHARMACEUTICAL AND BIOMEDICAL ANALYSIS, 1993, 11 (01) :1-13
[4]   A MULTIWAVELENGTH APPROACH TO THE SELECTION OF ABSORBENCY RATIOS FOR THE ASSESSMENT OF CHROMATOGRAPHIC PEAK PURITY [J].
CASTLEDINE, JB ;
FELL, AF ;
MODIN, R ;
SELLBERG, B .
JOURNAL OF PHARMACEUTICAL AND BIOMEDICAL ANALYSIS, 1991, 9 (08) :619-624
[5]   EVALUATION OF A PHOTODIODE ARRAY DETECTOR FOR THE VERIFICATION OF PEAK HOMOGENEITY IN HIGH-PERFORMANCE LIQUID-CHROMATOGRAPHY [J].
CHAN, HK ;
CARR, GP .
JOURNAL OF PHARMACEUTICAL AND BIOMEDICAL ANALYSIS, 1990, 8 (03) :271-277
[6]   LC DETECTORS - EVALUATION AND PRACTICAL IMPLICATIONS OF LINEARITY [J].
DORSCHEL, CA ;
EKMANIS, JL ;
OBERHOLTZER, JE ;
WARREN, FV ;
BIDLINGMEYER, BA .
ANALYTICAL CHEMISTRY, 1989, 61 (17) :A951-&
[7]   IMPLEMENTATION AND EFFICIENCY OF AN AUTOMATIC PEAK-PURITY-CONTROL PROCEDURE IN HPLC-UV-VIS-COUPLING BASED ON PRINCIPAL COMPONENT ANALYSIS [J].
EBEL, S ;
MUECK, W .
CHROMATOGRAPHIA, 1988, 25 (12) :1075-1086
[8]  
MENDER EI, 1987, J CHROMATOGR, V419, P135
[9]   MULTICHANNEL DETECTION IN HIGH-PERFORMANCE LIQUID-CHROMATOGRAPHY [J].
MILLER, JC ;
GEORGE, SA ;
WILLIS, BG .
SCIENCE, 1982, 218 (4569) :241-246
[10]  
RICHARDSON H, 1992, 1992 PITTSB C EXP NE