THE CONESPRAY NEBULIZER FOR INDUCTIVELY COUPLED PLASMA ATOMIC EMISSION-SPECTROMETRY

被引:23
作者
IVALDI, JC [1 ]
VOLLMER, J [1 ]
SLAVIN, W [1 ]
机构
[1] PERKIN ELMER CORP,761 MAIN AVE,NORWALK,CT 06859
关键词
PNEUMATIC NEBULIZERS; SIGNAL; SPRAY; FLOW;
D O I
10.1016/0584-8547(91)80102-9
中图分类号
O433 [光谱学];
学科分类号
0703 ; 070302 ;
摘要
A version of the conespray nebulizer proposed by Sharp has been designed and evaluated for routine liquid sample introduction for inductively coupled plasma atomic emission spectrometry (ICP-AES). The short-term analytical precision was better than 1% RSD with the conespray nebulizer using both simple aqueous solutions and solutions containing a matrix of 10% NaCl. The long-term %RSD (about 8 hr) was typically about 1% with simple (low matrix) solutions. The precision and stability of the analytical signal while nebulizing solutions with a 10% NaCl matrix during continuous runs were not distinguishable from the case when simple solutions were used. Sensitivity and detection limit data collected for the conespray nebulizer at a solution flow rate of 2 ml/min and the cross-flow nebulizer at 1 ml/min showed similar performance for these figures of merit. Wash-in and wash-out times for the conespray nebulizer were 9 and 40 s, respectively. Generally, this version of the conespray nebulizer provided similar overall performance to the best of the currently available Babington style nebulizers such as the V-groove nebulizer.
引用
收藏
页码:1063 / 1072
页数:10
相关论文
共 18 条
[1]   INTERFERENCE FROM EASILY IONIZABLE ELEMENT MATRICES IN INDUCTIVELY COUPLED PLASMA EMISSION-SPECTROMETRY - A SPATIAL STUDY [J].
BLADES, MW ;
HORLICK, G .
SPECTROCHIMICA ACTA PART B-ATOMIC SPECTROSCOPY, 1981, 36 (09) :881-900
[2]  
BOUMANS PWJ, IN PRESS SPECTROCHIM, V46
[4]   EVALUATION OF THE GRID-TYPE NEBULIZER FOR THE INTRODUCTION OF HIGH DISSOLVED SALT AND HIGH SOLIDS CONTENT SOLUTIONS INTO THE INDUCTIVELY COUPLED PLASMA [J].
BROTHERTON, T ;
CARUSO, J .
JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY, 1987, 2 (07) :695-698
[5]   A STUDY OF PNEUMATIC NEBULIZATION SYSTEMS FOR INDUCTIVELY COUPLED PLASMA EMISSION-SPECTROMETRY [J].
EBDON, L ;
CAVE, MR .
ANALYST, 1982, 107 (1271) :172-178
[6]   INFLUENCE OF SOLUTION UPTAKE RATE ON SIGNALS AND INTERFERENCES IN INDUCTIVELY COUPLED PLASMA OPTICAL-EMISSION SPECTROMETRY [J].
FARINO, J ;
MILLER, JR ;
SMITH, DD ;
BROWNER, RF .
ANALYTICAL CHEMISTRY, 1987, 59 (18) :2303-2309
[7]   CROSS-FLOW NEBULIZERS AND TESTING PROCEDURES FOR INDUCTIVELY COUPLED PLASMA NEBULIZERS [J].
IVALDI, JC ;
SLAVIN, W .
JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY, 1990, 5 (05) :359-364
[8]   PRESERVATION OF ACCURACY AND PRECISION IN THE ANALYTICAL PRACTICE OF LOW-POWER ICP-AES [J].
MAESSEN, FJMJ ;
BALKE, J ;
DEBOER, JLM .
SPECTROCHIMICA ACTA PART B-ATOMIC SPECTROSCOPY, 1982, 37 (06) :517-526
[9]   COMPARISON OF PNEUMATIC NEBULIZERS IN CURRENT USE FOR INDUCTIVELY COUPLED PLASMA ATOMIC EMISSION-SPECTROMETRY [J].
MAESSEN, FJMJ ;
COEVERT, P ;
BALKE, J .
ANALYTICAL CHEMISTRY, 1984, 56 (06) :899-903
[10]   SOME EFFECTS OF LOW ACID CONCENTRATIONS IN INDUCTIVELY COUPLED PLASMA ATOMIC EMISSION-SPECTROMETRY [J].
MARICHY, M ;
MERMET, M ;
MERMET, JM .
SPECTROCHIMICA ACTA PART B-ATOMIC SPECTROSCOPY, 1990, 45 (11) :1195-1201