IMPROVEMENT OF SIGNAL-TO-NOISE RATIO OF PHOTOMULTIPLIERS FOR VERY WEAK SIGNALS

被引:35
作者
TOPP, JA
SCHROTTER, HW
HACKER, H
BRANDMULLER, J
机构
[1] Sektion Physik der Universität München
关键词
D O I
10.1063/1.1684189
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
For the photoelectric detection of Raman spectra excited with a He-Ne laser, the dark current of an EMI 9558 QA photomultiplier is reduced by magnetic defocusing of the electrons originating from the inefficient part of the photocathode. A lightpipe cone concentrates the light signal to a spot of 2.7 mm diameter and enhances the quantum efficiency of the photocathode by multiple reflection. A pulse counting system with a deadtime of less than 100 nsec is described and compared with dc detection. It is now possible to detect signals with a power of less than 10-17 W or less than 35 photons/sec at 650 nm with a time constant of 1 sec. The improvement in signal-to-noise ratio is demonstrated with the Raman spectrum of polycrystalline azobenzene. © 1969 The American Institute of Physics.
引用