TISSUE TEMPERATURE MONITORING FOR THERMAL INTERVENTIONAL THERAPY - COMPARISON OF T1-WEIGHTED MR SEQUENCES

被引:86
作者
MATSUMOTO, R
MULKERN, RV
HUSHEK, SG
JOLESZ, FA
机构
[1] HARVARD UNIV,BRIGHAM & WOMENS HOSP,SCH MED,DEPT RADIOL,75 FRANCIS ST,BOSTON,MA 02115
[2] CHILDRENS HOSP MED CTR,DEPT RADIOL,BOSTON,MA 02115
[3] MIT,DEPT NUCL ENGN,CAMBRIDGE,MA 02139
来源
JMRI-JOURNAL OF MAGNETIC RESONANCE IMAGING | 1994年 / 4卷 / 01期
关键词
PULSE SEQUENCES; RAPID IMAGING; TEMPERATURE MONITORING;
D O I
10.1002/jmri.1880040114
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
For thermal interventional therapy, near real-time monitoring of temperature changes in the treated area is desirable. In this study, various fast T1-weighted magnetic resonance (MR) imaging protocols were compared to determine the sensitivity and resolution of signal intensity for temperatures within the range of 36-degrees-C-66-degrees-C in gel phantoms and in vitro porcine liver specimens. The results showed that a T1-weighted fast spin-echo sequence with a TR of 100 msec had better temperature sensitivity and resolution than other sequences with comparable temporal resolutions. The longer imaging times required for fast spin-echo sequences with a TR of 300 msec did not improve temperature sensitivity. The methods introduced to evaluate temperature sensitivity and resolution should prove useful in selecting appropriate MR protocols for monitoring thermal treatment modalities such as interstitial laser therapy, focused ultrasound therapy, or radio-frequency heating.
引用
收藏
页码:65 / 70
页数:6
相关论文
共 13 条
[1]   THE POLYACRYLAMIDE AS A PHANTOM MATERIAL FOR ELECTROMAGNETIC HYPERTHERMIA STUDIES [J].
BINI, MG ;
IGNESTI, A ;
MILLANTA, L ;
OLMI, R ;
RUBINO, N ;
VANNI, R .
IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 1984, 31 (03) :317-322
[2]   REAL-TIME MAGNETIC-RESONANCE-IMAGING OF LASER HEAT DEPOSITION IN TISSUE [J].
BLEIER, AR ;
JOLESZ, FA ;
COHEN, MS ;
WEISSKOFF, RM ;
DALCANTON, JJ ;
HIGUCHI, N ;
FEINBERG, DA ;
ROSEN, BR ;
MCKINSTRY, RC ;
HUSHEK, SG .
MAGNETIC RESONANCE IN MEDICINE, 1991, 21 (01) :132-137
[3]   MR-GUIDED FOCUSED ULTRASOUND SURGERY [J].
CLINE, HE ;
SCHENCK, JF ;
HYNYNEN, K ;
WATKINS, RD ;
SOUZA, SP ;
JOLESZ, FA .
JOURNAL OF COMPUTER ASSISTED TOMOGRAPHY, 1992, 16 (06) :956-965
[4]  
GEWIESE B, 1992, BOOK ABSTRACTS SOC M, P793
[5]   RARE IMAGING - A FAST IMAGING METHOD FOR CLINICAL MR [J].
HENNIG, J ;
NAUERTH, A ;
FRIEDBURG, H .
MAGNETIC RESONANCE IN MEDICINE, 1986, 3 (06) :823-833
[6]   MR IMAGING OF LASER-TISSUE INTERACTIONS [J].
JOLESZ, FA ;
BLEIER, AR ;
JAKAB, P ;
RUENZEL, PW ;
HUTTL, K ;
JAKO, GJ .
RADIOLOGY, 1988, 168 (01) :249-253
[7]   TEMPERATURE MAPPING WITH MR IMAGING OF MOLECULAR-DIFFUSION - APPLICATION TO HYPERTHERMIA [J].
LEBIHAN, D ;
DELANNOY, J ;
LEVIN, RL .
RADIOLOGY, 1989, 171 (03) :853-857
[8]  
LEWA CJ, 1980, B CANCER, V67, P525
[9]   MONITORING OF LASER AND FREEZING-INDUCED ABLATION IN THE LIVER WITH T1-WEIGHTED MR IMAGING [J].
MATSUMOTO, R ;
OSHIO, K ;
JOLESZ, FA .
JMRI-JOURNAL OF MAGNETIC RESONANCE IMAGING, 1992, 2 (05) :555-562
[10]   PARTIAL RF ECHO-PLANAR IMAGING WITH THE FAISE METHOD .2. CONTRAST EQUIVALENCE WITH SPIN-ECHO SEQUENCES [J].
MELKI, PS ;
JOLESZ, FA ;
MULKERN, RV .
MAGNETIC RESONANCE IN MEDICINE, 1992, 26 (02) :342-354