PROXIMITY FOCUSED HYBRID PHOTO DIODE CHARACTERISTICS EVALUATIONS

被引:26
作者
ARNAUDON, H
BEMNETTI, P
BOSKMA, L
BURGER, P
CONFIN, A
DESALVO, R
PAOLUCCI, S
ROSSELLA, M
SCHOMAKER, R
机构
[1] CERN, LAA, CH-1211 GENEVA 23, SWITZERLAND
[2] UNIV STRASBOURG 1, F-67070 STRASBOURG, FRANCE
[3] INFN, SEZ PAVIA, I-27100 PAVIA, ITALY
[4] DEP, 9300 AB RODEN, NETHERLANDS
[5] CANBERRA, B-2250 OLEN, BELGIUM
[6] INFN, SEZ BOLOGNA, I-40126 BOLOGNA, ITALY
[7] UNIV BOLOGNA, I-40126 BOLOGNA, ITALY
关键词
D O I
10.1016/0168-9002(94)90287-9
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
The Hybrid PhotoDiode (HPD) tube is a new kind of photodetector. Characteristics of proximity focussed HPD tube prototypes are reported. Insensitivity to magnetic fields to the tesla level, fast response (few ns rise and fall time), good response uniformity and linearity over 8 orders of magnitude were measured. Particular attention was paid to voltage and charge response to high intensity light pulses. The HPD prototypes reported here are the precursors of a family of multipixel HPDs foreseen for the second half of 1993.
引用
收藏
页码:558 / 570
页数:13
相关论文
共 9 条
[1]   TEST-RESULTS OF THE 1ST PROXIMITY FOCUSED HYBRID PHOTODIODE DETECTOR PROTOTYPES [J].
BASA, S ;
CLEMENS, JC ;
COMMERCON, M ;
SAUVAGE, D ;
TISSERANT, S ;
GAILLARD, JM .
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 1993, 330 (1-2) :93-99
[2]  
CUSHMAN P, 1992, IN PRESS 4TH ANN IIS
[3]   1ST RESULTS ON THE HYBRID PHOTODIODE TUBE [J].
DESALVO, R ;
HAO, W ;
YOU, K ;
WANG, Y ;
XU, C .
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 1992, 315 (1-3) :375-384
[4]  
DESALVO R, 1987, CLNS8792 CORN U PREP
[5]   OPERATIONAL CHARACTERISTICS OF AN ELECTRON-BOMBARDED SILICON-DIODE PHOTOMULTIPLIER TUBE [J].
JOHANSEN, GA ;
JOHNSON, CB .
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 1993, 326 (1-2) :295-298
[6]   SUBNANOSECOND TIMING WITH ION-IMPLANTED DETECTORS [J].
RIJKEN, HA ;
KLEIN, SS ;
JACOBS, W ;
TEEUWEN, LJHGW ;
DEVOIGT, MJA ;
BURGER, P .
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION B-BEAM INTERACTIONS WITH MATERIALS AND ATOMS, 1992, 64 (1-4) :272-276
[7]  
SCLAR N, 1957, ELECTRON DEVICES C W
[8]  
STOOP KWJ, 1991, VIDEOGRAPHY PHOTONIC, V1539, P138
[9]   HYBRID PHOTOTUBE WITH SI TARGET [J].
VANGEEST, LK ;
STOOP, KWJ .
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 1991, 310 (1-2) :261-266