NONINVASIVE DIAGNOSIS OF CORONARY-ARTERY DISEASE USING A NEURAL NETWORK ALGORITHM

被引:47
作者
AKAY, M
机构
[1] Biomedical Engineering Department, Rutgers University, College of Engineering, Piscataway, 08855-0909, NJ
关键词
D O I
10.1007/BF02414891
中图分类号
TP3 [计算技术、计算机技术];
学科分类号
0812 ;
摘要
This study examines the utility of neural networks for detecting coronary artery disease noninvasively by using the clinical examination variables and extracting useful information from the diastolic heart sounds associated with coronary occlusions. It has been widely reported that coronary stenoses produce sounds due to the turbulent blood flow in these vessels. These complex and highly attenuated signals taken from recordings made in both soundproof and noisy rooms were detected and analyzed to provide feature set based on the poles and power spectral density function (PSD) of the Autoregressive (AR) method after Adaptive Line Enhancement (ALE) method. In addition, some physical examination variables such as sex, age, body weight, smoking condition, diastolic pressure, systolic pressure and derivation from them were included in the feature vector. This feature vector was used as the input pattern to the neural network. The analysis was studied on one hundred recordings (63 abnormal, 37 normals). The network correctly identified 84% of the subjects with coronary artery disease and 89% of the normal subjects.
引用
收藏
页码:361 / 367
页数:7
相关论文
共 58 条
  • [1] APPLICATION OF THE ARMA METHOD TO ACOUSTIC DETECTION OF CORONARY-ARTERY DISEASE
    AKAY, M
    WELKOWITZ, W
    SEMMLOW, JL
    KOSTIS, J
    [J]. MEDICAL & BIOLOGICAL ENGINEERING & COMPUTING, 1991, 29 (04) : 365 - 372
  • [2] NONINVASIVE DETECTION OF CORONARY STENOSES BEFORE AND AFTER ANGIOPLASTY USING EIGENVECTOR METHODS
    AKAY, M
    SEMMLOW, JL
    WELKOWITZ, W
    BAUER, MD
    KOSTIS, JB
    [J]. IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 1990, 37 (11) : 1095 - 1104
  • [3] DETECTION OF CORONARY OCCLUSIONS USING AUTOREGRESSIVE MODELING OF DIASTOLIC HEART SOUNDS
    AKAY, M
    SEMMLOW, JL
    WELKOWITZ, W
    BAUER, MD
    KOSTIS, JB
    [J]. IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 1990, 37 (04) : 366 - 373
  • [4] AKAY M, 1990, THESIS RUTGERS U PIS
  • [5] AKAY M, 1992, IN PRESS MED BIOL EN
  • [6] BEAT-TO-BEAT DETECTION OF HIS-PURKINJE SYSTEM SIGNALS USING ADAPTIVE FILTERS
    ALNASHASH, HAM
    KELLY, SW
    TAYLOR, DJE
    [J]. MEDICAL & BIOLOGICAL ENGINEERING & COMPUTING, 1988, 26 (02) : 117 - 125
  • [7] Berne R.M., 1986, CARDIOVASCULAR PHYSL
  • [8] BAYESIAN STATISTICS AS APPLIED TO HYPERTENSION DIAGNOSIS
    BLINOWSKA, A
    CHATELLIER, G
    BERNIER, J
    LAVRIL, M
    [J]. IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 1991, 38 (07) : 699 - 706
  • [9] ADAPTIVE METHOD FOR CANCELLATION OF RESPIRATORY ARTIFACT IN ELECTROGASTRIC MEASUREMENTS
    CHEN, JD
    VANDEWALLE, J
    SANSEN, W
    VANTRAPPEN, G
    JANSSENS, J
    [J]. MEDICAL & BIOLOGICAL ENGINEERING & COMPUTING, 1989, 27 (01) : 57 - 63
  • [10] DIASTOLIC MURMUR CAUSED BY CORONARY ARTERY STENOSIS
    CHENG, TO
    [J]. ANNALS OF INTERNAL MEDICINE, 1970, 72 (04) : 543 - +