CORROSION AND WEAR AT THE MODULAR INTERFACE OF UNCEMENTED FEMORAL STEMS

被引:91
作者
COOK, SD
BARRACK, RL
CLEMOW, AJT
机构
来源
JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME | 1994年 / 76B卷 / 01期
关键词
D O I
10.1302/0301-620X.76B1.8300685
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
We examined 108 uncemented femoral stems with modular femoral heads which had been retrieved for reasons other than loosening. There were detectable amounts of wear and corrosion in 10 of 29 (34.5%) mixed-alloy components and 7 of 79 (9%) single-alloy components after a mean implantation time of 25 months. We found no correlation between the presence or extent of corrosion or surface damage and any of time in situ, initial diagnosis, reason for removal, age, or weight. Stems with wear and corrosion were less likely to show histological bony ingrowth. The interface between the head and stem of modular total hip components is a possible source of ion release and wear debris, but wear and corrosion were totally absent in most specimens. This suggests that this problem could be avoided, and that further research is required to develop manufacturing methods which would minimise such changes.
引用
收藏
页码:68 / 72
页数:5
相关论文
共 21 条
  • [1] BAUER TW, 1992, 38TH T ORTH RES SOC, P354
  • [2] CORROSION BETWEEN THE COMPONENTS OF MODULAR FEMORAL HIP PROSTHESES
    COLLIER, JP
    SURPRENANT, VA
    JENSEN, RE
    MAYOR, MB
    SURPRENANT, HP
    [J]. JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME, 1992, 74 (04): : 511 - 517
  • [3] COLLIER JP, 1992, 38TH P ANN M ORTH RE, P355
  • [4] COLLIER JP, 1991, CLIN ORTHOP RELAT R, V271, P305
  • [5] FRETTING CORROSION IN ORTHOPEDIC ALLOYS
    COOK, SD
    GIANOLI, GJ
    CLEMOW, AJT
    HADDAD, RJ
    [J]. BIOMATERIALS MEDICAL DEVICES AND ARTIFICIAL ORGANS, 1983, 11 (04): : 281 - 292
  • [6] COOK SD, 1991, P COMBINED M ORTHOPA, P229
  • [7] COOK SD, 1991, J ARTHROPLASTY S, V6, P37
  • [8] DEVINE TM, 1974, THESIS MIT
  • [9] FRETTING CORROSION STUDIES OF UNIVERSAL FEMORAL-HEAD PROSTHESES AND CONE TAPER SPIGOTS
    FRICKER, DC
    SHIVANATH, R
    [J]. BIOMATERIALS, 1990, 11 (07) : 495 - 500
  • [10] HUO MH, 1991, ORTHOPEDICS, V14, P335