A SIMPLE DIFFERENTIAL THERMOMETRIC TECHNIQUE FOR MEASURING RATES OF CHEMICAL REACTIONS IN SOLUTIONS

被引:29
作者
MEITES, T
MEITES, L
JAITLY, JN
机构
[1] Department of Chemistry, Clarkson College of Technology, Potsdam
关键词
D O I
10.1021/j100845a039
中图分类号
O64 [物理化学(理论化学)、化学物理学];
学科分类号
070304 ; 081704 ;
摘要
This paper describes the construction and operation of a constant-temperature-environment differential twin microcalorimetric apparatus for following the rate of a chemical reaction by monitoring the difference between the temperature of a reaction mixture, contained in one of two identical vessels, with that of a reference solution contained in the other. The temperatures are sensed by thermistors, whose resistances are compared in a simple Wheatstone bridge. Data on the reaction between ethyl acetate and sodium hydroxide indicate that second-order rate constants can be evaluated with accuracies and precisions of a few per cent or better even if the overall change of temperature during a reaction is as small as 0.015°. Various tests for internal consistency indicate that the measured differential temperatures are reliable to ± 20 μdeg or better.
引用
收藏
页码:3801 / &
相关论文
共 45 条
[41]  
SWINBOURNE ES, 1960, J CHEM SOC, P2371
[42]   REACTION RATE OF ALKALINE HYDROLYSIS OF ETHYL ACETATE [J].
TSUJIKAWA, H ;
INOUE, H .
BULLETIN OF THE CHEMICAL SOCIETY OF JAPAN, 1966, 39 (09) :1837-+
[43]  
TYRRELL HJV, 1968, THERMOMETRIC TITRIME, P40
[44]  
Wolf KL, 1940, Z PHYS CHEM A-CHEM T, V187, P61
[45]   A THERMAL METHOD FOR MEASURING THE RATES OF MODERATELY FAST REACTIONS [J].
WYATT, PAH .
JOURNAL OF THE CHEMICAL SOCIETY, 1960, (MAY) :2299-2302