PULSED CARBON-DIOXIDE LASER ABLATION OF BURNED SKIN - INVITRO AND INVIVO ANALYSIS

被引:59
作者
GREEN, HA
DOMANKEVITZ, Y
NISHIOKA, NS
机构
[1] Wellman Laboratory of Biomedical Laser Applications, Department of Dermatology, Harvard Medical School, Massachusetts General Hospital, Boston, Massachusetts
关键词
excision; pig skin; rat skin; surgery;
D O I
10.1002/lsm.1900100513
中图分类号
R75 [皮肤病学与性病学];
学科分类号
100206 ;
摘要
Pulsed lasers produce efficient and precise tissue ablation with limited residual thermal damage. In this study, the efficiency of pulsed CO2 laser ablation of burned and normal swine skin was studied in vitro with a mass loss technique. The heats of ablation for normal and burned skin were 2,706 and 2,416 J/cm3 of tissue ablated, respectively. The mean threshold radiant exposures for ablating normal skin and eschar were 2.6 J/cm2 and 3.0 J/cm2, respectively. Radiant exposures greater than 19 J/cm2 produced a plasma, which decreased the efficiency of laser ablation. Thus the radiant exposures for efficient ablation range from 4 to 19 J/cm2, and within this radiant exposure range 20–40 μm of tissue are ablated per pulse. We also examined, on a gross and histo‐pathologic basis, in vivo burn eschar excision with a pulsed CO2 laser. The laser allowed bloodless excisions of full thickness burns on the backs of male hairless rats. The zone of thermal damage was approximately 85 μm over the subjacent fascia. The pulsed CO2 laser can ablate burn eschar efficiently, precisely, and bloodlessly and may prove valuable for the excision of burned and necrotic tissue. Copyright © 1990 Wiley‐Liss, Inc., A Wiley Company
引用
收藏
页码:476 / 484
页数:9
相关论文
共 48 条
[41]   BURN ESCHAR - HISTOPATHOLOGICAL STUDY [J].
VISTNES, LM ;
HOGG, GR .
PLASTIC AND RECONSTRUCTIVE SURGERY, 1971, 48 (01) :56-&
[42]   PULSED CO2-LASER TISSUE ABLATION - EFFECT OF TISSUE-TYPE AND PULSE DURATION ON THERMAL-DAMAGE [J].
WALSH, JT ;
FLOTTE, TJ ;
ANDERSON, RR ;
DEUTSCH, TF .
LASERS IN SURGERY AND MEDICINE, 1988, 8 (02) :108-118
[43]   ER YAG LASER ABLATION OF TISSUE - EFFECT OF PULSE DURATION AND TISSUE-TYPE ON THERMAL-DAMAGE [J].
WALSH, JT ;
FLOTTE, TJ ;
DEUTSCH, TF .
LASERS IN SURGERY AND MEDICINE, 1989, 9 (04) :314-326
[44]   PULSED CO2-LASER TISSUE ABLATION - MEASUREMENT OF THE ABLATION RATE [J].
WALSH, JT ;
DEUTSCH, TF .
LASERS IN SURGERY AND MEDICINE, 1988, 8 (03) :264-275
[45]  
WARNER RR, 1986, MICROBEAM ANAL, V21, P238
[46]  
WEAST RC, 1987, CRC HDB CHEM PHYSICS
[47]  
ZOLTAREV VM, 1969, OPT SPEKTROSK+, V27, P430
[48]   MECHANICAL AND THERMAL PARAMETERS IN PULSED LASER CUTTING OF TISSUE [J].
ZWEIG, AD ;
WEBER, HP .
IEEE JOURNAL OF QUANTUM ELECTRONICS, 1987, 23 (10) :1787-1793