FABRICATION AND CHARACTERIZATION OF A SCREEN-PRINTED, DISPOSABLE, AMPEROMETRIC CHOLESTEROL BIOSENSOR

被引:62
作者
GILMARTIN, MAT [1 ]
HART, JP [1 ]
机构
[1] UNIV W ENGLAND, FAC SCI APPL, ELECTROCHEM SENSORS & BIOSENSORS GRP, BRISTOL BS16 1QY, AVON, ENGLAND
关键词
CHOLESTEROL BIOSENSOR; AMPEROMETRY; SCREEN PRINTING; ELECTROCATALYSIS; PERMSELECTIVITY;
D O I
10.1039/an9941902331
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
A single-use, reagentless, screen-printed biosensor strip for the measurement of free cholesterol was fabricated and its performance characteristics were evaluated by cyclic voltammetry and under steady-state conditions. Cholesterol oxidase was adsorbed on the surface of an H2O2-sensing cobalt phthalocyanine screen-printed carbon electrode and retained at the sensory interface with a solvent-east cellulose acetate membrane. The device, operated in the amperometric mode, provides an inexpensive, reliable and rapid means of quantifying this clinically significant sterol. Its analytical utility is further augmented by its ease of fabrication and favourable storage properties. The biosensor exhibits a wide functional range (6 x 10(-5)-5.0 x 10(-3) mol dm(-3)) and the sensitivities of the devices (slopes of the calibration graphs) were used to derive optimum solution conditions. Several analytical parameters were also investigated to optimize and enhance the response time and sensitivity of the sensor and these are also discussed.
引用
收藏
页码:2331 / 2336
页数:6
相关论文
共 31 条
[1]  
ALLAIN CC, 1974, CLIN CHEM, V20, P470
[2]   AMPEROMETRIC ASSAY FOR THE KETONE-BODY 3-HYDROXYBUTYRATE [J].
BATCHELOR, MJ ;
GREEN, MJ ;
SKETCH, CL .
ANALYTICA CHIMICA ACTA, 1989, 221 (02) :289-294
[3]   THE DETERMINATION OF TOTAL SERUM-CHOLESTEROL BY FLOW-INJECTION ANALYSIS WITH AMPEROMETRIC DETECTION [J].
CARPENTER, A ;
PURDY, WC .
ANALYTICAL LETTERS, 1990, 23 (03) :425-435
[4]   CHOLESTEROL SENSOR-BASED ON ELECTRODEPOSITION OF CATALYTIC PALLADIUM PARTICLES [J].
DONG, SJ ;
DENG, Q ;
CHENG, GG .
ANALYTICA CHIMICA ACTA, 1993, 279 (02) :235-240
[5]  
FLEGG H M, 1973, Annals of Clinical Biochemistry, V10, P79, DOI 10.1177/000456327301000125
[6]  
FREDRICKSON DS, 1972, METABOLIC BASIS INHE, P545
[7]   DEVELOPMENT OF AMPEROMETRIC SENSORS FOR URIC-ACID BASED ON CHEMICALLY-MODIFIED GRAPHITE-EPOXY RESIN AND SCREEN-PRINTED ELECTRODES CONTAINING COBALT PHTHALOCYANINE [J].
GILMARTIN, MAT ;
HART, JP ;
BIRCH, BJ .
ANALYST, 1994, 119 (02) :243-252
[8]   VOLTAMMETRIC AND AMPEROMETRIC BEHAVIOR OF URIC-ACID AT BARE AND SURFACE-MODIFIED SCREEN-PRINTED ELECTRODES - STUDIES TOWARDS A DISPOSABLE URIC-ACID SENSOR [J].
GILMARTIN, MAT ;
HART, JP .
ANALYST, 1992, 117 (08) :1299-1303
[9]   NOVEL, REAGENTLESS, AMPEROMETRIC BIOSENSOR FOR URIC-ACID BASED ON A CHEMICALLY-MODIFIED SCREEN-PRINTED CARBON ELECTRODE COATED WITH CELLULOSE-ACETATE AND URICASE [J].
GILMARTIN, MAT ;
HART, JP .
ANALYST, 1994, 119 (05) :833-840
[10]  
GILMARTIN MAT, UNPUB ANALYST