FOCAL MAGNETIC STIMULATION OF AN AXON

被引:47
作者
BASSER, PJ
机构
[1] Biomedical Engineering and Instrumentation Program, National Center of Research Resources, National Institutes of Health, Bethesda
关键词
D O I
10.1109/10.293248
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
The induced electric field produced by a circular coil during magnetic stimulation of an axon is derived from Maxwell's equations. The foci and virtual cathodal and anodal regions are predicted as a function of coil radius and orientation. Two virtual anode and cathode pairs are predicted, one lying outside the coil's perimeter and predominant in the far field? and one lying within the perimeter of the coil which may stimulate the axon when the coil and nerve are in close proximity. When the coil is positioned tangent to the nerve, an orientation commonly used in clinical magnetic stimulation, the foci of the predominant cathode acid anode pair are extremely sensitive to changes tn coil placement. In addition, the radius of curvature of the activating function, a measure of the size of the virtual cathode at threshold, is predicted to decrease with decreasing coil diameter and distance to the nerve. These predictions may help explain observed variability in measurements of conduction velocity and latency during magnetic stimulation of peripheral axons.
引用
收藏
页码:601 / 606
页数:6
相关论文
共 24 条
[11]   THE UTILITY OF MAGNETIC STIMULATION FOR ROUTINE PERIPHERAL-NERVE CONDUCTION STUDIES [J].
EVANS, BA ;
LITCHY, WJ ;
DAUBE, JR .
MUSCLE & NERVE, 1988, 11 (10) :1074-1078
[12]   MAGNETISM - A NEW METHOD FOR STIMULATION OF NERVE AND BRAIN [J].
HALLETT, M ;
COHEN, LG .
JAMA-JOURNAL OF THE AMERICAN MEDICAL ASSOCIATION, 1989, 262 (04) :538-541
[13]  
MACCABBEE PJ, 1990, 20TH P ANN M SOC NEU, P1261
[14]   MAGNETIC COIL STIMULATION OF STRAIGHT AND BENT AMPHIBIAN AND MAMMALIAN PERIPHERAL-NERVE INVITRO - LOCUS OF EXCITATION [J].
MACCABEE, PJ ;
AMASSIAN, VE ;
EBERLE, LP ;
CRACCO, RQ .
JOURNAL OF PHYSIOLOGY-LONDON, 1993, 460 :201-219
[15]   DETERMINING THE SITE OF STIMULATION DURING MAGNETIC STIMULATION OF A PERIPHERAL-NERVE [J].
NILSSON, J ;
PANIZZA, M ;
ROTH, BJ ;
BASSER, PJ ;
COHEN, LG ;
CARUSO, G ;
HALLETT, M .
ELECTROENCEPHALOGRAPHY AND CLINICAL NEUROPHYSIOLOGY, 1992, 85 (04) :253-264
[16]  
NILSSON J, 1991, 43RD P ANN M AM AC N
[17]   ANALYSIS OF MODELS FOR EXTERNAL STIMULATION OF AXONS [J].
RATTAY, F .
IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 1986, 33 (10) :974-977
[18]   A THEORETICAL CALCULATION OF THE ELECTRIC-FIELD INDUCED BY MAGNETIC STIMULATION OF A PERIPHERAL-NERVE [J].
ROTH, BJ ;
COHEN, LG ;
HALLETT, M ;
FRIAUF, W ;
BASSER, PJ .
MUSCLE & NERVE, 1990, 13 (08) :734-741
[19]   A MODEL OF THE STIMULATION OF A NERVE-FIBER BY ELECTROMAGNETIC INDUCTION [J].
ROTH, BJ ;
BASSER, PJ .
IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 1990, 37 (06) :588-597
[20]   IN-VITRO EVALUATION OF A 4-LEAF COIL DESIGN FOR MAGNETIC STIMULATION OF PERIPHERAL-NERVE [J].
ROTH, BJ ;
MACCABEE, PJ ;
EBERLE, LP ;
AMASSIAN, VE ;
HALLETT, M ;
CADWELL, J ;
ANSELMI, GD ;
TATARIAN, GT .
ELECTROENCEPHALOGRAPHY AND CLINICAL NEUROPHYSIOLOGY, 1994, 93 (01) :68-74