INTENSITY-MODULATED ARC THERAPY WITH DYNAMIC MULTILEAF COLLIMATION - AN ALTERNATIVE TO TOMOTHERAPY

被引:518
作者
YU, CX
机构
[1] William Beaumont Hospital, Royal Oak, MI
关键词
D O I
10.1088/0031-9155/40/9/004
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
The desire to improve local tumour control and cure more cancer patients, coupled with advances in computer technology and linear accelerator design, has spurred the developments of three-dimensional conformal radiotherapy techniques. Optimized treatment plans, aiming to deliver high dose to the target while minimizing dose to the surrounding tissues, can be delivered with multiple fields each with spatially modulated beam intensities or with multiple-slice treatments. This paper introduces a new method, intensity-modulated are therapy (IMAT), for delivering optimized treatment plans to improve the therapeutic ratio. It utilizes continuous gantry motion as in conventional are therapy. Unlike conventional are therapy, the field shape, which is conformed with the multileaf collimator, changes during gantry rotation. Arbitrary two-dimensional beam intensity distributions at different beam angles are delivered with multiple superimposing arcs. A system capable of delivering the IMAT has been implemented. An example is given that illustrates the feasibility of this new method. Advantages of this new technique over tomotherapy and other slice-based delivery schemes are also discussed.
引用
收藏
页码:1435 / 1449
页数:15
相关论文
共 25 条
[1]   METHODS OF IMAGE-RECONSTRUCTION FROM PROJECTIONS APPLIED TO CONFORMATION RADIOTHERAPY [J].
BORTFELD, T ;
BURKELBACH, J ;
BOESECKE, R ;
SCHLEGEL, W .
PHYSICS IN MEDICINE AND BIOLOGY, 1990, 35 (10) :1423-1434
[2]   X-RAY FIELD COMPENSATION WITH MULTILEAF COLLIMATORS [J].
BORTFELD, TR ;
KAHLER, DL ;
WALDRON, TJ ;
BOYER, AL .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 1994, 28 (03) :723-730
[3]  
BOYER AL, 1994, 11TH P INT C US COMP, P178
[4]   OPTIMIZATION OF STATIONARY AND MOVING BEAM RADIATION-THERAPY TECHNIQUES [J].
BRAHME, A .
RADIOTHERAPY AND ONCOLOGY, 1988, 12 (02) :129-140
[5]   OPTIMIZATION OF RADIATION-THERAPY [J].
BRAHME, A .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 1994, 28 (03) :785-787
[6]  
CAROL MP, 1994, 11 INT C US COMP RAD, P172
[7]  
CAROL MP, 1993, 3 DIMENSIONAL TREATM, P173
[8]   THE GENERATION OF INTENSITY-MODULATED FIELDS FOR CONFORMAL RADIOTHERAPY BY DYNAMIC COLLIMATION [J].
CONVERY, DJ ;
ROSENBLOOM, ME .
PHYSICS IN MEDICINE AND BIOLOGY, 1992, 37 (06) :1359-1374
[9]   THE EFFECT OF LOCAL-CONTROL ON METASTATIC DISSEMINATION IN CARCINOMA OF THE PROSTATE - LONG-TERM RESULTS IN PATIENTS TREATED WITH I-125 IMPLANTATION [J].
FUKS, Z ;
LEIBEL, SA ;
WALLNER, KE ;
BEGG, CB ;
FAIR, WR ;
ANDERSON, LL ;
HILARIS, BS ;
WHITMORE, WF .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 1991, 21 (03) :537-547
[10]   A GENERALIZED PENCIL BEAM ALGORITHM FOR OPTIMIZATION OF RADIATION-THERAPY [J].
GUSTAFSSON, A ;
LIND, BK ;
BRAHME, A .
MEDICAL PHYSICS, 1994, 21 (03) :343-357