PRECISE POSITIONING OF INTRACRANIAL SMALL TUMORS TO THE LINEAR ACCELERATORS ISOCENTER, USING A STEREOTAXIC RADIOTHERAPY COMPUTED-TOMOGRAPHY SYSTEM (SRT-CT)

被引:15
作者
SHIRATO, H
SUZUKI, K
NISHIOKA, T
KAMADA, T
KAGEI, K
KITAHARA, T
MORISAWA, H
TSUJII, H
机构
[1] OBIHIRO KOHSEI HOSP,DEPT THERAPEUT & DIAGNOST RADIOL,OBIHIRO,HOKKAIDO,JAPAN
[2] SHIMADZU CO LTD,DIV MED SYST,KYOTO,JAPAN
[3] PROTON MED RES CTR,TSUKUBA,IBARAKI,JAPAN
关键词
RADIOTHERAPY; 3-DIMENSIONAL; TREATMENT PLANNING; SIMULATOR; COMPUTED TOMOGRAPHY;
D O I
10.1016/0167-8140(94)90105-8
中图分类号
R73 [肿瘤学];
学科分类号
100214 ;
摘要
RT-CT was developed as a simulator using CT scan for radiotherapy. Following three-dimensional treatment planning using CT images, the treatment center and treatment fields are projected to the patients' surface by laser beam on the C arm. The prototype had an accuracy of 3 mm which was equivalent to conventional X-ray simulators but was not adequate for use in stereotactic radiotherapy. A new RT-CT system was developed to have a precise localization capability for stereotactic radiotherapy. Using this stereotactic RT-CT (SRT-CT) after three-dimensional planning, the treatment center is projected to the stereotactic frame automatically. In this study, the values of the x, y and z coordinates of the target center determined by SRT-CT are compared with those determined by the traditional method using CT localizing plates. The discrepancies were within 1.0 mm in 90% and 1.5 mm in 100% of 30 measurements in 16 patients. The disadvantages of SRT-CT may be that the accuracy of localization depends on the quality of calibration of laser beams. The traditional CT localizing method has superiority over SRT-CT because of its solid coordinates but its accuracy is vulnerable to alignment of CT fiducial marker plates. Therefore, the SRT-CT and traditional CT localizing methods would be complementary to each other for precise localization.
引用
收藏
页码:180 / 183
页数:4
相关论文
共 9 条
[1]  
HENRI CJ, 1991, MED PHYS, V18, P168
[2]  
IRIE G, 1980, 7TH P ICCR INT C US, P355
[3]   CT SIMULATOR - A NEW 3-D PLANNING AND SIMULATING SYSTEM FOR RADIOTHERAPY .2. CLINICAL-APPLICATION [J].
NAGATA, Y ;
NISHIDAI, T ;
ABE, M ;
TAKAHASHI, M ;
OKAJIMA, K ;
YAMAOKA, N ;
ISHIHARA, H ;
KUBO, Y ;
OHTA, H ;
KAZUSA, C .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 1990, 18 (03) :505-513
[4]   CT SIMULATOR - A NEW 3-D PLANNING AND SIMULATING SYSTEM FOR RADIOTHERAPY .1. DESCRIPTION OF SYSTEM [J].
NISHIDAI, T ;
NAGATA, Y ;
TAKAHASHI, M ;
ABE, M ;
YAMAOKA, N ;
ISHIHARA, H ;
KUBO, Y ;
OHTA, H ;
KAZUSA, C .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 1990, 18 (03) :499-504
[5]  
PHILLIPS MH, 1993, PHYSICAL ASPECTS STE, P1
[6]   RADIOSURGERY TARGET POINT ALIGNMENT ERRORS DETECTED WITH PORTAL FILM VERIFICATION [J].
SERAGO, CF ;
LEWIN, AA ;
HOUDEK, PV ;
GONZALEZARIAS, S ;
SCHWADE, JG ;
ABITOL, A ;
MARCIALVEGA, V .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 1992, 24 (04) :777-780
[7]  
SHIRATO H, 1992, J JPN SOC THER RADIO, V4, P149
[8]   QUALITY ASSURANCE IN STEREOTAXIC RADIOSURGERY USING A STANDARD LINEAR-ACCELERATOR [J].
TSAI, JS ;
BUCK, BA ;
SVENSSON, GK ;
ALEXANDER, E ;
CHENG, CW ;
MANNARINO, EG ;
LOEFFLER, JS .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 1991, 21 (03) :737-748
[9]  
TSUJII H, 1987, 9TH P ICCR INT C US, P185