MEASUREMENT POSSIBILITIES USING AN ELECTRONIC PORTAL IMAGING DEVICE

被引:63
作者
KIRBY, MC
WILLIAMS, PC
机构
[1] North Western Medical Physics Department, Christie Hospital, Manchester
关键词
PORTAL IMAGING; EXIT DOSIMETRY; QUALITY CONTROL; MULTILEAF COLLIMATOR;
D O I
10.1016/0167-8140(93)90253-5
中图分类号
R73 [肿瘤学];
学科分类号
100214 ;
摘要
A vital role in the quality control of radiotherapy is the use of portal imaging for verifying field size, shape, orientation and patient set-up. Coincidence of treated volume and target volume is imperative. Electronic portal imaging devices are effective at providing this verification. However, these devices are versatile enough to be used in other ways pertinent to the delivery of high quality, high precision radiotherapy. This paper examines two such ways: in assessing the reproducibility of a multileaf collimator system, and in determining exit doses in vivo. Configured as a dosemeter, the system shows a linear response with good dynamic ranger Its high spatial resolution was used to show that leaf positioning was reproducible to within 0.5 mm for all tested gantry and collimator angles. Our preliminary results from this exit dosimetry technique demonstrate that, under specific conditions, doses can be determined to within 2.5% of that measured using silicon diodes or ion chambers.
引用
收藏
页码:237 / 243
页数:7
相关论文
共 39 条
  • [1] AMOLS HI, 1987, MED PHYS, V14, P483
  • [2] BAILY NA, 1980, INT J RADIAT ONCOL, V6, P935, DOI 10.1016/0360-3016(80)90341-7
  • [3] RADIATION-FIELD EDGE-DETECTION IN PORTAL IMAGES
    BIJHOLD, J
    GILHUIJS, KGA
    VANHERK, M
    MEERTENS, H
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 1991, 36 (12) : 1705 - 1710
  • [4] A REVIEW OF ELECTRONIC PORTAL IMAGING DEVICES (EPIDS)
    BOYER, AL
    ANTONUK, L
    FENSTER, A
    VANHERK, M
    MEERTENS, H
    MUNRO, P
    REINSTEIN, LE
    WONG, J
    [J]. MEDICAL PHYSICS, 1992, 19 (01) : 1 - 16
  • [5] ROUTINE CLINICAL ONLINE PORTAL IMAGING FOLLOWED BY IMMEDIATE FIELD ADJUSTMENT USING A TELE-CONTROLLED PATIENT COUCH
    DENEVE, W
    VANDENHEUVEL, F
    DEBEUKELEER, M
    COGHE, M
    THON, L
    DEROOVER, P
    VANLANCKER, M
    STORME, G
    [J]. RADIOTHERAPY AND ONCOLOGY, 1992, 24 (01) : 45 - 54
  • [6] IMAGE COMPARISON TECHNIQUES FOR USE WITH MEGAVOLTAGE IMAGING-SYSTEMS
    EVANS, PM
    GILDERSLEVE, JQ
    MORTON, EJ
    SWINDELL, W
    COLES, R
    FERRARO, M
    RAWLINGS, C
    XIAO, ZR
    DYER, J
    [J]. BRITISH JOURNAL OF RADIOLOGY, 1992, 65 (776) : 701 - 709
  • [7] AUTOMATIC ONLINE INSPECTION OF PATIENT SETUP IN RADIATION-THERAPY USING DIGITAL PORTAL IMAGES
    GILHUIJS, KGA
    VANHERK, M
    [J]. MEDICAL PHYSICS, 1993, 20 (03) : 667 - 677
  • [8] A METHOD TO ANALYZE 2-DIMENSIONAL DAILY RADIOTHERAPY PORTAL IMAGES FROM AN ONLINE FIBEROPTIC IMAGING-SYSTEM
    GRAHAM, ML
    CHENG, AY
    GEER, LY
    BINNS, WR
    VANNIER, MW
    WONG, JW
    [J]. INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 1991, 20 (03): : 613 - 619
  • [9] STUDY OF TREATMENT VARIATION IN THE RADIOTHERAPY OF HEAD AND NECK TUMORS USING A FIBEROPTIC ONLINE RADIOTHERAPY IMAGING-SYSTEM
    HALVERSON, KJ
    LEUNG, TC
    PELLET, JB
    GERBER, RL
    WEINHOUS, MS
    WONG, JW
    [J]. INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 1991, 21 (05): : 1327 - 1336
  • [10] KINZIE JJ, 1983, CANCER, V52, P2223, DOI 10.1002/1097-0142(19831215)52:12<2223::AID-CNCR2820521209>3.0.CO