REVIEW OF CONVERGENT BEAM TOMOGRAPHY IN SINGLE PHOTON-EMISSION COMPUTED-TOMOGRAPHY

被引:87
作者
GULLBERG, GT
ZENG, GL
DATZ, FL
CHRISTIAN, PE
TUNG, CH
MORGAN, HT
机构
[1] UNIV UTAH,DEPT RADIOL,DIV NUCL MED,SALT LAKE CITY,UT 84132
[2] OHIO IMAGING PICKER INT,BEDFORD HTS,OH 44128
关键词
D O I
10.1088/0031-9155/37/3/002
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Investigation of convergent-beam single photon emission computed tomography (SPECT) is actively being pursued to evaluate its clinical potential. Fan-beam, cone-beam, pin-hole and astigmatic collimators are being used with rotating gamma cameras having large crystal areas, to increase the sensitivity for emission and transmission computed tomography of small organs such as the thyroid, brain or heart. With new multi-detector SPECT systems, convergent-beam geometry offers the ability to simultaneously obtain emission and transmission data necessary to quantify uptake of radiopharmaceutical distributions in the heart. The development of convergent-beam geometry in SPECT requires the integration of hardware and software. In considering hardware, the optimum detector system for cone-beam tomography is a system that satisfies the data sufficiency condition for which the scanning trajectory intersects any plane passing through the reconstructed region of interest. However, the major development of algorithms has been for the data insufficient case of single planar orbit acquisitions. The development of these algorithms have made possible the preliminary evaluation of this technology and the imaging of brain and heart are showing significant potential for the clinical application of cone-beam tomography. Presently, significant research activity is pursuing the development of algorithms for data acquisitions that satisfy the data sufficiency condition and that can be implemented easily and inexpensively on clinical SPECT systems.
引用
收藏
页码:507 / 534
页数:28
相关论文
共 55 条
[41]  
TSUI BMW, 1989, J NUCL MED, V30, P497
[42]   IMPLEMENTATION OF SIMULTANEOUS ATTENUATION AND DETECTOR RESPONSE CORRECTION IN SPECT [J].
TSUI, BMW ;
HU, HB ;
GILLAND, DR ;
GULLBERG, GT .
IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 1988, 35 (01) :778-783
[43]  
TSUI BMW, 1989, J NUCL MED, V30, P882
[44]  
TUNG C H, 1991, Journal of Nuclear Medicine, V32, P955
[45]  
TUNG CH, 1991, J NCUL MED, V323, P1066
[46]   AN INVERSION-FORMULA FOR CONE-BEAM RECONSTRUCTION [J].
TUY, HK .
SIAM JOURNAL ON APPLIED MATHEMATICS, 1983, 43 (03) :546-552
[47]   A COMPARISON OF ATTENUATION CORRECTION METHODS FOR QUANTITATIVE SINGLE PHOTON-EMISSION COMPUTED-TOMOGRAPHY [J].
WEBB, S ;
FLOWER, MA ;
OTT, RJ ;
LEACH, MO .
PHYSICS IN MEDICINE AND BIOLOGY, 1983, 28 (09) :1045-1056
[48]   FORMATION OF IMAGES USING FAN-BEAM SCANNING AND NONCIRCULAR SOURCE MOTION [J].
WEINSTEIN, FS .
JOURNAL OF THE OPTICAL SOCIETY OF AMERICA, 1980, 70 (08) :931-935
[49]  
ZENG G L, 1990, Journal of Nuclear Medicine, V31, P867
[50]   3-DIMENSIONAL ITERATIVE RECONSTRUCTION ALGORITHMS WITH ATTENUATION AND GEOMETRIC POINT RESPONSE CORRECTION [J].
ZENG, GL ;
GULLBERG, GT ;
TSUI, BMW ;
TERRY, JA .