VERIFICATION OF GAMMA-SPECTROSCOPY PROGRAMS - MULTIPLET AREA PROBLEMS AND SOLUTIONS

被引:9
作者
KOSKELO, MJ [1 ]
MERCIER, MT [1 ]
机构
[1] CANBERRA NUCL PROD GRP, ITASCA, IL 60413 USA
来源
JOURNAL OF RADIOANALYTICAL AND NUCLEAR CHEMISTRY-ARTICLES | 1995年 / 193卷 / 02期
关键词
D O I
10.1007/BF02039877
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
In this work we demonstrate that the latest addition to the Canberra/Nuclear Data family of software packages, Genie-PC, performs as well as or better than the older software packages with respect to the doubler tests reported on earlier. In addition, we demonstrate the behavior of the various software packages for a doublet which includes the 511 keV annihilation line. And we show that using other means to break a doubler into its component areas will also lead to accurate overall results.