BIOCOMPATIBILITY OF HEPARIN-COATED MEMBRANE-OXYGENATOR DURING CARDIOPULMONARY BYPASS

被引:22
作者
HATORI, N
YOSHIZU, H
HAGA, Y
KUSAMA, Y
TAKESHIMA, S
SEGAWA, D
TANAKA, S
机构
[1] Department of Surgery II, National Defense Medical College, Saitama
关键词
HEPARIN-COATED OXYGENATOR; CORONARY ARTERY BYPASS; CARDIOPULMONARY BYPASS; PLATELET; GRANULOCYTE; COMPLEMENT; PROSTAGLANDIN;
D O I
10.1111/j.1525-1594.1994.tb03342.x
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
The biocompatibility of the cardiopulmonary bypass (CPB) circuit, in which an oxygenator is solely heparinized, was assessed by systemic inflammatory reactions as an indicator during CPB. Fourteen patients, 11 males and 3 females, underwent coronary artery bypass surgery and were randomly divided into 2 groups of 7 patients each. For the heparin-coated oxygenator group (Group H), a heparin-coated membrane oxygenator was used in the CPB circuit, and in the control (Group C) an uncoated membrane oxygenator was employed. Systemic inflammatory reactions, such as platelet activation, prostaglandin production, complement activation, and activated granulocyte released substance, were measured prior to, during, and 6 h after CPB. The number of platelets decreased after protamine administration in both groups (14.5 +/- 4.7 x 10(4)/mu l in Group H and 13.8 +/- 8.7 x 10(4)/mu l in Group C) and returned to baseline levels in Group H while it remained decreased in Group C at 6 h after CPB. The platelet factor 4 level was significantly lower in Group H (181 +/- 40 ng/ml) than in Group C (297 +/- 131 ng/ml) after protamine administration. Thromboxane-B-2 (TXB(2)) rose during CPB in both groups; however, there were significantly different levels of TXB(2) between the 2 groups at 60 min after CPB (293 +/- 258 pg/ml in Group H versus 408 +/- 120 pg/ml in Group C) and after protamine administration (259 +/- 122 pg/ml in Group H versus 709 +/- 418 pg/ml in Group C). Plasma concentrations of granulocyte elastase were significantly lower in Group H at 30, 60 and 90 min, immediately after, and post-CPB than those of Group C. Although the oxygenator was solely heparinized in the CPB circuit, it was sufficiently effective to reduce inflammatory reactions during coronary artery bypass operation, and the heparin-coated surface seems to be more endothelium-like.
引用
收藏
页码:904 / 910
页数:7
相关论文
共 37 条
[21]  
KIRKLIN JK, 1983, J THORAC CARDIOV SUR, V86, P845
[22]  
KIRKLIN JW, 1980, MAYO CLIN PROC, V55, P339
[23]   ACTIVATION OF INFLAMMATORY SYSTEMS DURING CARDIOPULMONARY BYPASS [J].
NILSSON, L ;
BRUNNKVIST, S ;
NILSSON, U ;
NYSTROM, SO ;
TYDEN, H ;
VENGE, P ;
ABERG, T .
SCANDINAVIAN JOURNAL OF THORACIC AND CARDIOVASCULAR SURGERY, 1988, 22 (01) :51-53
[24]  
PASCHE B, 1986, THROMB RES, P44
[25]   NEUTROPHIL DE-GRANULATION AND INCREASED PULMONARY CAPILLARY-PERMEABILITY FOLLOWING ESOPHAGECTOMY - A MODEL OF EARLY LUNG INJURY IN MAN [J].
ROCKER, GM ;
WISEMAN, MS ;
PEARSON, D ;
SHALE, DJ .
BRITISH JOURNAL OF SURGERY, 1988, 75 (09) :883-886
[26]   COMPLEMENT ACTIVATION IN CARDIOPULMONARY BYPASS, WITH SPECIAL REFERENCE TO ANAPHYLATOXIN PRODUCTION IN MEMBRANE AND BUBBLE OXYGENATORS [J].
TAMIYA, T ;
YAMASAKI, M ;
MAEO, Y ;
YAMASHIRO, T ;
OGOSHI, S ;
FUJIMOTO, S .
ANNALS OF THORACIC SURGERY, 1988, 46 (01) :47-57
[27]   COMPLEMENT ACTIVATION AND LUNG PERMEABILITY DURING CARDIOPULMONARY BYPASS [J].
TENNENBERG, SD ;
CLARDY, CW ;
BAILEY, WW ;
SOLOMKIN, JS .
ANNALS OF THORACIC SURGERY, 1990, 50 (04) :597-601
[28]   HEPARIN-COATED CARDIOPULMONARY BYPASS CIRCUITS REDUCE BLOOD-CELL TRAUMA - EXPERIMENTS IN THE PIG [J].
THELIN, S ;
BAGGE, L ;
HULTMAN, J ;
BOROWIEC, J ;
NILSSON, L ;
THORELIUS, J .
EUROPEAN JOURNAL OF CARDIO-THORACIC SURGERY, 1991, 5 (09) :486-491
[29]  
VANOEVEREN W, 1985, J THORAC CARDIOV SUR, V89, P888
[30]  
VIDEM V, 1991, J THORAC CARDIOV SUR, V101, P654