RESOLUTION CORRECTION FOR SURFACE X-RAY-DIFFRACTION AT HIGH BEAM EXIT ANGLES

被引:21
作者
SCHAMPER, C
MEYERHEIM, HL
MORITZ, W
机构
[1] Inst fur Kristallographie and, Mineralogie der Universitat Munchen, Munchen
关键词
D O I
10.1107/S0021889893004364
中图分类号
O6 [化学];
学科分类号
0703 ;
摘要
Owing to the two-dimensional periodicity of a super-structure on the crystal surface, the intensity in reciprocal space is continuously distributed along rods normal to the sample surface. The analysis of rod scans in surface X-ray diffraction provides information about the structure parameters normal to the sample surface. For high resolution to be achieved, the measurements must extend to momentum transfers q(perpendicular-to) that are as large as possible. At large exit angles, the conventional Lorentz factor must be modified to take account of the finite aperture of the detector and the continuous intensity along the lattice rod. For two types of Z-axis diffractometer used in surface X-ray crystallography, an analytical expression for the resolution correction of rod-scan intensity data has been developed. It takes into account an anisotropic detector resolution T(DELTATHETA, DELTAPHI), the finite width of the diffracted beam and the primary-beam divergence parallel to the sample surface, DELTA(tau). The calculation of the convolution functions is simplified by a projection onto the q(perpendicular-to) = 0 plane. The effects of different detector settings and the influences of the primary-beam divergence and the sample quality on the measured intensity are demonstrated for several examples.
引用
收藏
页码:687 / 696
页数:10
相关论文
共 14 条
[1]   SCATTERING OF X-RAYS FROM CRYSTAL-SURFACES [J].
ANDREWS, SR ;
COWLEY, RA .
JOURNAL OF PHYSICS C-SOLID STATE PHYSICS, 1985, 18 (35) :6427-6439
[2]   ANGLE AND INDEX CALCULATIONS FOR A Z-AXIS X-RAY DIFFRACTOMETER [J].
BLOCH, JM .
JOURNAL OF APPLIED CRYSTALLOGRAPHY, 1985, 18 (FEB) :33-36
[3]   RESOLUTION CORRECTIONS IN DIFFUSE-SCATTERING EXPERIMENTS [J].
BOYSEN, H ;
ADLHART, W .
JOURNAL OF APPLIED CRYSTALLOGRAPHY, 1987, 20 :200-209
[4]   A NOVEL X-RAY-SCATTERING DIFFRACTOMETER FOR STUDYING SURFACE-STRUCTURES UNDER UHV CONDITIONS [J].
BRENNAN, S ;
EISENBERGER, P .
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 1984, 222 (1-2) :164-167
[5]   SURFACE-STRUCTURE DETERMINATION BY X-RAY-DIFFRACTION [J].
FEIDENHANSL, R .
SURFACE SCIENCE REPORTS, 1989, 10 (03) :105-188
[6]  
KOCH EE, 1983, HDB SYNCHROTRON RAD, V1, P1
[7]  
KRETSCHMAR F, 1987, Z KRISTALLOGR, V178, P130
[8]  
Robinson I. K., 1991, HDB SYNCHROTRON RAD, V3, P221
[9]   THE GE(001) (2 X-1) RECONSTRUCTION - ASYMMETRIC DIMERS AND MULTILAYER RELAXATION OBSERVED BY GRAZING-INCIDENCE X-RAY-DIFFRACTION [J].
ROSSMANN, R ;
MEYERHEIM, HL ;
JAHNS, V ;
WEVER, J ;
MORITZ, W ;
WOLF, D ;
DORNISCH, D ;
SCHULZ, H .
SURFACE SCIENCE, 1992, 279 (1-2) :199-209
[10]   A new precision x-ray spectrometer [J].
Soller, W .
PHYSICAL REVIEW, 1924, 24 (02) :158-167