EVALUATION OF THE FACTORS AFFECTING THE ACCURACY AND PRECISION OF A TECHNIQUE FOR QUANTIFICATION OF VOLUME AND ACTIVITY IN SPECT

被引:15
作者
ALAAMER, AS
FLEMING, JS
PERRING, S
机构
[1] SOUTHAMPTON GEN HOSP,DEPT MED PHYS,SOUTHAMPTON SO9 4XY,HANTS,ENGLAND
[2] SOUTHAMPTON GEN HOSP,DEPT NUCL MED,SOUTHAMPTON SO9 4XY,HANTS,ENGLAND
关键词
D O I
10.1097/00006231-199409000-00016
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
The use of the technique of maximization of interclass variance to determine the threshold level defining the volume of an object in single photon emission computed tomography (SPECT) and hence its activity has been evaluated in a variety of conditions. The influence of varying activity on volume assessment and varying volume on activity assessment were shown to be negligible. The effect of object shape for the range of shapes likely to be found in the body was also small. Non-uniformity of activity in the object caused an underestimation of volume and to a lesser extent its activity. However, with activity varying by 100% within the volume, volume was only underestimated by 10% and activity by 3%. The presence of activity surrounding the object caused an overestimation of volume which increased linearly with the relative level of surrounding activity. The object to surrounding activity concentration ratio could be estimated and therefore its effect corrected. The influence of nearby objects and methods of dealing with their effect on the technique are described. Activity concentrations in objects with uniform distribution assessed using total activity divided by total volume were shown to be more accurate than assessments from maximum voxel count rate for objects with a volume below 150 ml.
引用
收藏
页码:758 / 771
页数:14
相关论文
共 21 条
[1]   EVALUATION OF A TECHNIQUE FOR THE QUANTIFICATION OF RADIOACTIVITY AND VOLUME OF AN OBJECT USING SPECT [J].
ALAAMER, AS ;
FLEMING, JS ;
PERRING, S .
NUCLEAR MEDICINE COMMUNICATIONS, 1993, 14 (12) :1061-1070
[2]  
ALAAMER AS, 1992, NUCL MED COMMUN, V13, P239
[3]  
CALDWELL JH, 1982, J NUCL MED, V23, P490
[5]   A TECHNIQUE FOR THE ABSOLUTE QUANTIFICATION OF I-131 RADIOPHARMACEUTICALS USING SPECT [J].
FLEMING, JS ;
ALAAMER, AA ;
PERRING, S .
NUCLEAR MEDICINE COMMUNICATIONS, 1993, 14 (06) :498-505
[6]   A GENERAL SOFTWARE PACKAGE FOR THE HANDLING OF MEDICAL IMAGES [J].
FLEMING, JS ;
BRITTEN, AJ ;
PERRING, S ;
KEEN, AC ;
HOWLETT, PJ .
JOURNAL OF MEDICAL ENGINEERING & TECHNOLOGY, 1991, 15 (4-5) :162-169
[7]   A 3-DIMENSIONAL 2ND-DERIVATIVE SURFACE-DETECTION ALGORITHM FOR VOLUME DETERMINATION ON SPECT IMAGES [J].
HILLEL, PG ;
HASTINGS, DL .
PHYSICS IN MEDICINE AND BIOLOGY, 1993, 38 (05) :583-600
[8]  
HILLEL PG, 1994, NUCL MED COMMUN, V15, P196
[9]  
IOSILEVSKY G, 1989, SEMIN NUCL MED, V19, P33
[10]   PHYSICAL FACTORS AFFECTING QUANTITATIVE MEASUREMENTS USING CAMERA-BASED SINGLE PHOTON-EMISSION COMPUTED-TOMOGRAPHY (SPECT) [J].
JASZCZAK, RJ ;
COLEMAN, RE ;
WHITEHEAD, FR .
IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 1981, 28 (01) :69-80