CONFOCAL THETA MICROSCOPE WITH 3 OBJECTIVE LENSES

被引:24
作者
LINDEK, S [1 ]
PICK, R [1 ]
STELZER, EHK [1 ]
机构
[1] RUPRECHT KARLS UNIV HEIDELBERG,INST ANGEW PHYS,D-69120 HEIDELBERG,GERMANY
关键词
D O I
10.1063/1.1144574
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
A microscope using three water immersion objective lenses which realizes confocal, 4Pi-confocal and various confocal theta microscopies in fluorescence, transmission, scattered, and reflection mode is described. An argon-ion laser is the primary light source. A pulsed titanium-sapphire laser allows two-photon absorption fluorescence microscopy. The instrument has a predicted resolution of 100 nm along the illumination axis and a three-dimensional resolution of 5×106 nm3 for lenses each with a numerical aperture of 0.75. This is an improvement of an order of magnitude over a confocal fluorescence microscope using the same lens. Applications of the microscope range from observation of a sample at three different angles, to confocal theta fluorescence microscopy with multiphoton absorption. Since mounting and immersion media are identical, aberrations become negligible. The large working distance of 2 mm makes the instrument ideal for the observation of biological samples of up to 1.5 mm in diameter such as drosophila embryos.
引用
收藏
页码:3367 / 3372
页数:6
相关论文
共 22 条
[1]  
[Anonymous], [No title captured], Patent No. [3013467 A, 3013467]
[2]  
[Anonymous], 1903, ANN PHYS, DOI DOI 10.1002/ANDP.19023150102
[3]   OPTICAL TRAPPING AND MANIPULATION OF SINGLE CELLS USING INFRARED-LASER BEAMS [J].
ASHKIN, A ;
DZIEDZIC, JM ;
YAMANE, T .
NATURE, 1987, 330 (6150) :769-771
[4]   A TILTING DEVICE FOR 3-DIMENSIONAL MICROSCOPY - APPLICATION TO INSITU IMAGING OF INTERPHASE CELL-NUCLEI [J].
BRADL, J ;
HAUSMANN, M ;
EHEMANN, V ;
KOMITOWSKI, D ;
CREMER, C .
JOURNAL OF MICROSCOPY, 1992, 168 :47-57
[5]   2-PHOTON LASER SCANNING FLUORESCENCE MICROSCOPY [J].
DENK, W ;
STRICKLER, JH ;
WEBB, WW .
SCIENCE, 1990, 248 (4951) :73-76
[6]  
FISCHER A, UNPUB APPL OPT
[7]   PROPERTIES OF A 4PI CONFOCAL FLUORESCENCE MICROSCOPE [J].
HELL, S ;
STELZER, EHK .
JOURNAL OF THE OPTICAL SOCIETY OF AMERICA A-OPTICS IMAGE SCIENCE AND VISION, 1992, 9 (12) :2159-2166
[8]   ABERRATIONS IN CONFOCAL FLUORESCENCE MICROSCOPY INDUCED BY MISMATCHES IN REFRACTIVE-INDEX [J].
HELL, S ;
REINER, G ;
CREMER, C ;
STELZER, EHK .
JOURNAL OF MICROSCOPY, 1993, 169 :391-405
[9]   MEASUREMENT OF THE 4PI-CONFOCAL POINT-SPREAD FUNCTION PROVES 75 NM AXIAL RESOLUTION [J].
HELL, SW ;
LINDEK, S ;
CREMER, C ;
STELZER, EHK .
APPLIED PHYSICS LETTERS, 1994, 64 (11) :1335-1337
[10]   ENHANCING THE AXIAL RESOLUTION IN FAR-FIELD LIGHT-MICROSCOPY - 2-PHOTON 4PI CONFOCAL FLUORESCENCE MICROSCOPY [J].
HELL, SW ;
LINDEK, S ;
STELZER, EHK .
JOURNAL OF MODERN OPTICS, 1994, 41 (04) :675-681