BACKSCATTERING AND TRANSMISSION THROUGH A HIGH-Z INTERFACE AS A MEASURE OF ELECTRON-BEAM ENERGY

被引:7
作者
DAS, IJ [1 ]
BUSHE, HS [1 ]
机构
[1] UNIV MASSACHUSETTS, MED CTR, DEPT RADIAT ONCOL, WORCESTER, MA 01655 USA
关键词
INTERFACE DOSIMETRY; ELECTRON BEAM; BACKSCATTER; TRANSMISSION; BEAM ENERGY; QUALITY ASSURANCE;
D O I
10.1118/1.597375
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
The ratio of electron beam transmission (Tr) and backscattering (Bs) is used as a measure of beam energy. With a proper thickness of high Z medium placed at the surface of a low Z homogeneous medium, it is observed that the function log(Tr/Bs) is a linear function of electron energy. This method is applied to electrons produced with scattering foil and pulse-Swept (scanned) beams for the nominal energies ranging from 6 to 20 MeV. This technique provides a unique alternative method which is easy, simple, and accurate within +/-3% for estimation of electron beam energy.